BREAST DOSES FROM PATIENTS AND FROM A STANDARD PHANTOM: ANALYSIS OF DIFFERENCES

机译:来自患者和标准体素的乳房剂量:差异分析

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The two methods commonly used for estimating the doses in mammography are based on the use of either a standard phantom or a sample of patients. Both methods are currently applied in the Galicia Breast Screening Programme. In the range of thicknesses corresponding to the standard-szied breast (50±5 mm), the use of a standard phantom resulted in an oversestimation of the ESAK values for patients by up 13/100. The overestimation depends on the view employed and is mroe important for the CC view. Differences in compostion between the standard breast and the real breasts are the main factor that affects the results obtained with both methods. Nevertheless, if a more realistic brast composition was considered, the ESAK vlaues from sample of patients would be clsoer to dose values obtained with the standard phantom.
机译:通常在乳房X线照相术中估计剂量的两种方法是基于标准体模或患者样本的使用。这两种方法目前都在加利西亚乳房筛查计划中应用。在对应于标准尺寸乳房的厚度范围内(50±5 mm),使用标准体模会导致患者对ESAK值的过度估计高达13/100。高估取决于所使用的视图,并且对于CC视图而言非常重要。标准乳房和真实乳房之间的成分差异是影响通过两种方法获得的结果的主要因素。但是,如果考虑使用更真实的胸罩组成,则来自患者样品的ESAK瓣将更接近使用标准体模获得的剂量值。