首页> 外文期刊>Radiation Protection Dosimetry >HOW DO HOSPITAL STERILISATION PROCEDURES AFFECT THE RESPONSE OF PERSONAL EXTREMITY RINGS AND OF EYE LENS TL DOSEMETERS?
【24h】

HOW DO HOSPITAL STERILISATION PROCEDURES AFFECT THE RESPONSE OF PERSONAL EXTREMITY RINGS AND OF EYE LENS TL DOSEMETERS?

机译:医院的消毒程序如何影响人的四肢和眼睛透镜TL剂量的反应?

获取原文
获取原文并翻译 | 示例
           

摘要

Stringent standards of hygiene must be applied in medical institutions, especially at operating blocks or during interventional radiology procedures. Medical equipment, including personal dosemeters that have to be worn by medical staff during such procedures, needs therefore to be sterilised. In this study, the effect of various sterilisation procedures has been tested on the dose response of extremity rings and of eye lens dosemeters in which thermoluminescent (TL) detectors (of types MTS-N and MCP-N, respectively) are used. The effects of medical sterilisation procedures were studied: by chemicals, by steam or by ultraviolet (UV), on the dose assessment by extremity rings and by eye lens dosemeters. Since it often happens that a dosemeter is accidentally machine-washed together with protective clothing, the effect of laundering on dose assessment by these dosemeters was also tested. The sterilisation by chemicals is mostly safe for TL detectors assuming that the dosemeters are waterproofed. Following sterilisation by water vapour, the response of these dosemeters diminished by some 30 %, irrespectively of the period of sterilisation; therefore, this method is not recommended. UV sterilisation can be applied to EYE-D™ eye lens dosemeters if their encapsulation is in black. The accidental dosemeter laundry in a washing machine has no impact on measured dose.
机译:医疗机构必须采用严格的卫生标准,尤其是在手术室或介入放射学过程中。因此,需要对医疗设备进行消毒,包括医疗人员在此类程序中必须佩戴的个人剂量计。在这项研究中,已经测试了各种灭菌程序对肢端环和使用热致发光(TL)检测器(分别为MTS-N和MCP-N型)的眼透镜剂量计的剂量响应的影响。研究了医疗灭菌程序的影响:通过化学药品,通过蒸汽或通过紫外线(UV),通过四肢环和眼镜镜片剂量计进行剂量评估。由于经常发生剂量计与防护服一起意外机洗的情况,因此还测试了洗涤对这些剂量计的剂量评估的影响。假设剂量计是防水的,则化学消毒对于TL检测器来说最安全。水蒸气灭菌后,这些剂量计的响应能力降低了约30%,与灭菌时间无关。因此,不建议使用此方法。如果封装形式为黑色,则可以对EYE-D™眼镜片剂量计进行紫外线消毒。洗衣机中的意外剂量计衣物对测量剂量没有影响。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2016年第4期|302-306|共5页
  • 作者单位

    Institute of Nuclear Physics Polish Academy of Science (IFJ PAN), ul. Radzikowskiego 152, PL 31-342 Krakow, Poland;

    Institute of Nuclear Physics Polish Academy of Science (IFJ PAN), ul. Radzikowskiego 152, PL 31-342 Krakow, Poland;

    Institute of Nuclear Physics Polish Academy of Science (IFJ PAN), ul. Radzikowskiego 152, PL 31-342 Krakow, Poland;

    Institute of Nuclear Physics Polish Academy of Science (IFJ PAN), ul. Radzikowskiego 152, PL 31-342 Krakow, Poland;

    Institute of Nuclear Physics Polish Academy of Science (IFJ PAN), ul. Radzikowskiego 152, PL 31-342 Krakow, Poland;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号