Waste reduction and process improvements in the analysis of plutonium by X-ray fluorescence: Results from multiple data sets

机译:X射线荧光分析reduction中的废物减少和工艺改进:来自多个数据集的结果

获取原文
获取原文并翻译 | 示例

摘要

To minimize waste, improve process safety, and reduce costs, modifications were implemented to a method for quantifying gallium in plutonium metal using wavelength dispersive X-ray fluorescence. These changes included reducing sample sizes, reducing ion exchange process volumes, using cheaper reagent grade acids, eliminating the use of HF acid, and using more robust containment films for sample analysis. Relative precision and accuracy achieved from analyzing multiple aliquots from a single parent sample were approximately 0.2 and 0.1%, respectively. The same precision was obtained from analyzing a total of four parent materials, and the average relative accuracy from all the samples was 0.4%, which is within programmatic uncertainty requirements.
机译:为了最大程度地减少浪费,提高过程安全性并降低成本,已对使用波长色散X射线荧光定量p金属中镓的方法进行了修改。这些变化包括减小样品尺寸,减少离子交换工艺量,使用更便宜的试剂级酸,取消使用HF酸以及使用更坚固的密封膜进行样品分析。通过分析来自单个母体样品的多个等分试样获得的相对精度和准确度分别约为0.2%和0.1%。通过分析总共四种母体材料获得相同的精度,并且所有样本的平均相对精度为0.4%,这在程序不确定性要求之内。