首页> 外文期刊>Microchimica Acta >Nonenzymatic hydrogen peroxide sensor based on a Prussian Blue-modified carbon ionic liquid electrode
【24h】

Nonenzymatic hydrogen peroxide sensor based on a Prussian Blue-modified carbon ionic liquid electrode

机译:基于普鲁士蓝修饰的碳离子液体电极的非酶过氧化氢传感器

获取原文
获取原文并翻译 | 示例
           

摘要

The electrocatalytic reduction of hydrogen peroxide (H2O2) was investigated using an electrode that was obtained by electropolymerization of Prussian Blue (PB) on a carbon ionic liquid electrode (CILE). The electrochemical reduction of H2O2 in buffer of pH 7.4 was performed using cyclic voltammetry and the amperometric I–t curve technique. A PB-modified CILE was applied to detect H2O2 at −0.10 V (versus the Ag/AgCl electrode). The linear dynamic range is from 0.05 to 6.0 mM, the detection limit is 1.0 µM (S/N = 3). The electrode displays good stability, a response time of <5 s, and better electrocatalytic activity towards H2O2. The ease and low cost of fabrication along with its high sensitivity and renewable surface make the CILE widely applicable in electroanalytical sensing. Keywords Carbon ionic liquid electrode - Prussian Blue - Electropolymerization
机译:使用在碳离子液体电极(CILE)上通过普鲁士蓝(PB)电聚合获得的电极研究了过氧化氢(H 2 O 2 )的电催化还原)。使用循环伏安法和安培I–t曲线技术对pH 7.4的缓冲液中H 2 O 2 进行电化学还原。 PB修饰的CILE用于检测-0.10 V(相对于Ag / AgCl电极)上的H 2 O 2 。线性动态范围为0.05到6.0 mM,检测极限为1.0 µM(S / N = 3)。该电极表现出良好的稳定性,响应时间<5 s,对H 2 O 2 具有更好的电催化活性。 CILE的简便性和低成本以及高灵敏度和可再生表面的特性使CILE广泛应用于电分析传感。碳离子液体电极普鲁士蓝电聚合