CALCULATION OF THE RADIATION DOSE ABSORBED BY HUMAN BIOLOGICAL TISSUE WHEN USING IMAGING EQUIPMENT IN RADIOTHERAPY

机译:在放射治疗中使用成像设备时,人体生物组织吸收的辐射剂量

获取原文
获取原文并翻译 | 示例
       

摘要

The issues of x-ray dose control during diagnostic and therapeutic procedures using imaging tools are considered. The dose of x-ray radiation absorbed by the biological tissue of a person from the visualization devices of the imaging procedure was determined while monitoring the position of the patient on the therapeutic table of the electron accelerator before the radiation therapy session. The processes of transmission of photons and electrons through the medium were simulated, and the x-ray spectra were measured. The emission spectrum of a Varian G-242 Rotating Anode X-Ray Tube was obtained using an XR-100-CdTe spectrometer. The absorbed dose was calculated by the Monte Carlo method. The dose absorbed in a water phantom at tube voltage up to 80 kV was 0.9-1.5 mGy.
机译:考虑了使用成像工具的诊断和治疗程序期间X射线剂量控制的问题。 通过从成像过程的可视化装置的人的生物组织吸收的X射线辐射的剂量,同时监测患者在放射治疗会议前的电子加速器的治疗表上的位置。 模拟光子和电子通过培养基的透射过程,测量X射线光谱。 使用XR-100-CDTE光谱仪获得Varian G-242旋转阳极X射线管的发射光谱。 通过蒙特卡罗方法计算吸收剂量。 在管电压下的水体模具中吸收的剂量,高达80kV为0.9-1.5毫米。