首页> 外文期刊>Lasers in Medical Science >A time resolved intense pulsed light spectral analysis system
【24h】

A time resolved intense pulsed light spectral analysis system

机译:时间分辨强脉冲光谱分析系统

获取原文
获取原文并翻译 | 示例
           

摘要

Intense pulsed light systems are currently in widespread use in clinical applications. Verification of spectral content and associated pulse waveforms is, however, not featured in product support and calibration. A measurement system was developed to determine the spectral output of such sources within a series of discrete wavelength intervals. Such a system utilised a series of 11 broadband optical filters with centre wavelengths within the range 450–950 nm in circuit with silicon photodiodes as optical detectors. Signals were captured using a low cost 8-channel 16-bit universal serial bus data capture module interfaced to a laptop computer. This allowed the pulse profile of selected filter components to be separately captured and analysed. Calibration of individual filter channels was undertaken using a Bentham dmc150 spectroradiometer and reference tungsten light source. Initial measurements undertaken of the intense pulsed light output of a Lumenis One system using the system are described.
机译:强脉冲光系统目前在临床应用中被广泛使用。但是,产品支持和校准中未提供光谱内容和相关脉冲波形的验证。开发了一种测量系统以确定一系列离散波长间隔内此类光源的光谱输出。这种系统在电路中使用了11个宽带滤光片,其中心波长在450-950 nm范围内,硅光电二极管作为光检测器。使用连接到便携式计算机的低成本8通道16位通用串行总线数据捕获模块捕获信号。这样可以分别捕获和分析所选滤波器组件的脉冲轮廓。使用Bentham dmc150分光光度计和参考钨光源对单个过滤器通道进行校准。描述了使用该系统对Lumenis One系统的强脉冲光输出进行的初始测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号