首页> 外文期刊>Journal of Hazardous Materials >Ultra-trace determination of silver in water samples by electrothermal atomic absorption spectrometry after preconcentration with a ligand-less cloud point extraction methodology
【24h】

Ultra-trace determination of silver in water samples by electrothermal atomic absorption spectrometry after preconcentration with a ligand-less cloud point extraction methodology

机译:无配体浊点萃取法预浓缩后,通过电热原子吸收光谱法超痕量测定水中的银

获取原文
获取原文并翻译 | 示例
       

摘要

A very simple and ligand-less cloud point extraction (CPE) methodology for the preconcentration of ultra-trace amounts of silver as a prior step to its determination by electrothermal atomic absorption spectrometry (ETAAS) has been developed. The method is based on the extraction of silver at pH 9 by using non-ionic surfactant polyethyleneglycolmono-p-nonylphenylether (PONPE 7.5) without adding any chelating agent. Several important variables that affect the CPE efficiency and ETAAS signal were investigated and optimized. The preconcentration of 15 ml sample solution allowed us to achieve an enhancement factor of 60. The calibration graph using the preconcentration system was linear in the range of 5-100 ng l~(-1) with a correlation coefficient of 0.9991. The lower limit of detection (3s) obtained in the optimal conditions was 1.2 ng l~(-1). The relative standard deviation (R.S.D.) for eight replicate determinations at 30 ng l~(-1) Ag level was 4.2%. The proposed method was successfully applied to the ultra-trace determination of silver in water samples.
机译:已经开发出一种非常简单且无配体的浊点萃取(CPE)方法,用于超痕量银的预浓缩,这是通过电热原子吸收光谱法(ETAAS)测定银的先前步骤。该方法基于在不添加任何螯合剂的情况下,通过使用非离子表面活性剂聚乙二醇单对壬基苯基醚(PONPE 7.5)提取pH为9的银的方法。研究和优化了影响CPE效率和ETAAS信号的几个重要变量。 15 ml样品溶液的预浓缩使我们获得了60的增强因子。使用该预浓缩系统的校准图在5-100 ng l〜(-1)范围内呈线性,相关系数为0.9991。在最佳条件下获得的检测下限(3s)为1.2 ng l〜(-1)。在30 ng l〜(-1)Ag浓度下进行八次重复测定的相对标准偏差(R.S.D.)为4.2%。该方法成功地用于水样中银的超痕量测定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号