首页> 外文期刊>Journal of Engineering in Medicine >A hip simulator study of metal-on-metal hip joint device using acetabular cups with different fixation surface conditions
【24h】

A hip simulator study of metal-on-metal hip joint device using acetabular cups with different fixation surface conditions

机译:使用髋臼杯固定表面条件不同的金属对金属髋关节装置的髋关节模拟器研究

获取原文
获取原文并翻译 | 示例
           

摘要

In vitro wear data for hip joint devices reported in the literature vary in a wide range from one simulator study to another sometimes for the same type of device tested under identical physiological testing conditions. We hypothesized that non-bearing surface condition of the testing components could be an important factor affecting the simulator wear results. To confirm this hypothesis, fifteen 50 mm metal-on-metal hip resurfacing devices with identical bearing specifications were tested in a ProSim hip wear simulator for 5 million cycles. The heads were standard Birmingham Hip Resurfacing (BHR) heads; whilst the pairing acetabular cups were identical to the standard BHR cup except their different back surface conditions, including: (a) off-the-shelf products after removing the hydroxyapatite (HA) coating; (b) semi-finished products without HA coating; and (c) purposely-made cups without cast-in beads and HA coating. Results showed that the different back surfaces of the cups used indeed caused significantly large variations in the gravimetrically measured wear loss. We postulated that materials loss from the non-bearing surface of the testing components could contribute to the gravimetrically measured wear loss during a wear simulator test both directly and indirectly. The results presented in this paper pertain to In vitro wear simulator study and have little clinical relevance to the performance of any implant in vivo.
机译:文献中报道的髋关节装置的体外磨损数据在很宽的范围内变化,从一次模拟器研究到另一次模拟器研究,有时是针对在相同生理测试条件下测试的同一类型的装置。我们假设测试组件的非轴承表面状况可能是影响模拟器磨损结果的重要因素。为了证实这一假设,在ProSim髋部磨损模拟器中对15个具有相同轴承规格的50 mm金属对金属髋关节表面重修设备进行了500万次循环测试。头是标准的伯明翰髋关节翻新(BHR)头。配对的髋臼杯除了背面条件不同外,其余均与标准BHR杯相同,包括:(a)去除羟基磷灰石(HA)涂层后的现成产品; (b)没有HA涂层的半成品; (c)特制的杯子,没有铸珠和HA涂层。结果表明,所用杯子的不同背面确实导致了重量分析的磨损损失的显着变化。我们假设,在磨损模拟器测试期间,直接或间接地,从测试组件的非承重表面产生的材料损失可能会导致重量测定的磨损损失。本文介绍的结果与体外穿戴模拟器的研究有关,与任何体内植入物的性能几乎没有临床相关性。

著录项

  • 来源
    《Journal of Engineering in Medicine》 |2011年第9期|p.877-887|共11页
  • 作者

  • 作者单位

    Implant Development Centre (IDC), Smith and Nephew Orthopaedics Ltd, Leamington Spa, UK;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号