A hip simulator study of metal-on-metal hip joint device using acetabular cups with different fixation surface conditions

机译:使用髋臼杯固定表面条件不同的金属对金属髋关节装置的髋关节模拟器研究

获取原文
获取原文并翻译 | 示例
           

摘要

In vitro wear data for hip joint devices reported in the literature vary in a wide range from one simulator study to another sometimes for the same type of device tested under identical physiological testing conditions. We hypothesized that non-bearing surface condition of the testing components could be an important factor affecting the simulator wear results. To confirm this hypothesis, fifteen 50mm metal-on-metal hip resurfacing devices with identical bearing specifications were tested in a ProSim hip wear simulator for 5 million cycles. The heads were standard Birmingham Hip Resurfacing (BHR) heads; whilst the pairing acetabular cups were identical to the standard BHR cup except their different back surface conditions, including: (a) off-the-shelf products after removing the hydroxyapatite (HA) coating; (b) semifinished products without HA coating; and (c) purposely-made cups without cast-in beads and HA coating. Results showed that the different back surfaces of the cups used indeed caused significantly large variations in the gravimetrically measured wear loss. We postulated that materials loss from the non-bearing surface of the testing components could contribute to the gravimetrically measured wear loss during a wear simulator test both directly and indirectly. The results presented in this paper pertain to In vitro wear simulator study and have little clinical relevance to the performance of any implant in vivo.
机译:文献中报道的髋关节器械的体外磨损数据变化很大,从一次模拟器研究到另一次模拟器研究,有时是针对在相同生理测试条件下测试的相同类型的器械。我们假设测试组件的非轴承表面状况可能是影响模拟器磨损结果的重要因素。为了证实这一假设,在ProSim髋部磨损模拟器中测试了15个具有相同轴承规格的50mm金属对金属髋关节表面重修设备,进行了500万次循环。头是标准的伯明翰髋关节翻新(BHR)头;配对的髋臼杯除了后背条件不同外,其余均与标准BHR杯相同,包括:(a)去除羟基磷灰石(HA)涂层后的现成产品; (b)没有HA涂层的半成品; (c)特制的无铸珠和HA涂层的杯子。结果表明,所使用的杯子的不同背面确实导致了重量分析的磨损损失的显着变化。我们假设,在磨损模拟器测试期间,通过重量分析测得的测试组件的非轴承表面的材料损耗可能直接或间接地导致了重量损失。本文介绍的结果与体外穿戴模拟器的研究有关,与任何体内植入物的性能均无临床相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号