首页> 外文期刊>Journal of British Podiatric Medicine >Mediclave or Medical Clave Steam Sterilisers - Inadequate Safety Locks
【24h】

Mediclave or Medical Clave Steam Sterilisers - Inadequate Safety Locks

机译:Mediclave或Medical Clave蒸汽灭菌器-安全锁不足

获取原文
获取原文并翻译 | 示例
       

摘要

The Medical Devices Agency has been informed by the Health and Safety Executive (HSE) that there are shortcomings in the design and construction of the safety interlocks on Mediclave benchtop steam sterilisers. The HSE has found the safety interlocks to be inadequate and not designed to be fail-safe. As a result it is possible to remove the boiler from the base, and to remove the boiler end cap, while the chamber contains steam under pressure. Mediclave benchtop steam sterilisers are not considered safe for use unless they are modified to ensure that while the chamber contains steam under pressure the boiler and cap cannot be removed.
机译:医疗设备局已从健康和安全执行官(HSE)获悉,Medicclave台式蒸汽灭菌器的安全互锁装置的设计和制造存在缺陷。 HSE发现安全联锁装置不足,并且没有设计成故障保险装置。结果,有可能在腔室中容纳有压力的蒸汽的同时,从底座上移除锅炉,并移除锅炉端盖。 Mediclave台式蒸汽灭菌器除非经过修改以确保在压力容器中盛装蒸汽的情况下不能移除锅炉和锅盖,否则认为它们不能安全使用。