Simultaneous adsorptive stripping voltammetric method for the determination of trace gallium(III) and germanium(IV) based on the adsorption of gallium(III) and germanium(IV)-catechol complex on the cyclic renewable mercury film silver based electrode (Hg(Ag)FE) is presented. The calibration graph is linear from 1.25 nM (0.09 μg L−1) to 90 nM (6.27 μg L−1) with correlation coefficient of 0.999 for gallium and from 2.5 nM (0.18 μg L−1) to 160 nM (12.3 μg L−1) with correlation coefficient of 0.998 for germanium for a preconcentration time of 30 s. The detection limit for a preconcentration time of 60 s is as low as 25 ng L−1 for gallium and 58 ng L−1 for germanium. The proposed method was successfully applied by studying the natural samples and simultaneous recovery of Ga(III) and Ge(IV) from spiked water and sediment samples.
机译:循环吸附式汞薄膜银基电极(Hg(Ag))吸附镓(III)和锗(IV)-邻苯二酚络合物的同时吸附溶出伏安法测定痕量镓(III)和锗(IV) FE)。校准图在1.25 nM(0.09μgL -1 )到90 nM(6.27μgL -1 )之间呈线性关系,镓的相关系数为0.999,2.5nM (0.18μgL -1 )到160 nM(12.3μgL -1 ),锗的预浓缩时间为30 s,相关系数为0.998。镓的预浓缩时间的检测限低至25 ng L -1 ,锗的低至58 ng L -1 。该方法通过研究天然样品并同时从加标的水和沉积物样品中同时回收Ga(III)和Ge(IV)而成功应用。