High-pitch dual-source CT coronary angiography: analysis of the impact on image quality of altered electrocardiography waves during data acquisition

机译:高螺距双源CT冠状动脉造影:分析数据采集过程中改变的心电图波对图像质量的影响

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Electrocardiography (ECG) “altered waves” sometimes occur during data acquisition when computed tomography coronary angiography (CTCA) is performed with the prospectively ECG-triggered high-pitch (Flash spiral) mode using a second-generation dual-source CT. The aim of this study was to assess the effect of the ECG altered waves on image quality. Seventy-three consecutive patients with stable sinus rhythm ≤65 beats per minute were retrospectively enrolled in this study. CTCA was performed using the Flash spiral mode in which the data acquisition was prospectively triggered at 60 % of the R–R interval and completed within one cardiac cycle. The ECG waves before and during data acquisition were analyzed for grouping purposes. Image quality was evaluated using a four-point scale (1 = best, 4 = unevaluatable). Thirty patients (group 1) were found to have ECG altered waves during data acquisition, while 43 patients (group 2) had ECG “stable waves.” The altered waves were seen as the baseline drifting; the broad, erected, or inverted P wave or QRS complexes; and a new wave. However, the length of the R–R interval did not change during the data acquisition. There were no significant differences in image quality scores between the two groups on the per-patient (2 ± 0.87 vs. 2.2 ± 0.74, P = 0.273) or per-segment (1.27 ± 0.54 vs. 1.32 ± 0.55, P = 0.577) basis. There were no significant differences in coronary evaluatability as well (per-patient; 93.3 vs. 95.3 %, P = 0.352; per-segment; 99.4 vs. 99.6 %, P = 1.0). CTCA image quality is not affected by ECG altered waves during data acquisition using the Flash spiral mode in low and stable heart rate patients. Thus, the ECG altered waves are considered artifacts.
机译:当使用前瞻性ECG触发的高音高(Flash螺旋)模式使用第二代双源CT进行计算机断层扫描冠状动脉造影(CTCA)时,数据采集期间有时会发生心电图(ECG)“改变的波”。这项研究的目的是评估心电图改变波对图像质量的影响。回顾性研究了73例连续窦性心律≤65次/分钟的连续患者。 CTCA是使用Flash螺旋模式执行的,该模式中的数据采集预期在R–R间隔的60%触发并在一个心动周期内完成。为了分组的目的,分析了数据采集之前和期间的ECG波。使用四点标度评估图像质量(1 =最佳,4 =不可评估)。发现30名患者(第1组)在数据采集过程中心电图改变,而43名患者(第2组)有心电图“稳定波”。改变后的波被视为基线漂移。宽,竖立或倒置的P波或QRS波群;和新的一波潮。但是,在数据采集过程中,R–R间隔的长度没有改变。两组在每位患者(2±0.87 vs. 2.2±0.74,P = 0.273)或每段(1.27±0.54 vs. 1.32±0.55,P = 0.577)之间的图像质量得分之间无显着差异基础。冠状动脉可评价性也无显着差异(每位患者; 93.3%vs. 95.3%,P = 0.352;每段; 99.4%vs. 99.6%,P = 1.0)。在心率稳定的低位患者中,使用Flash螺旋模式采集数据期间,CTCA图像质量不受心电图改变波的影响。因此,心电图改变的波被认为是伪像。

著录项

  • 来源
  • 作者单位

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Cardiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Guang An Men Hospital China Academy of Chinese Medical Sciences Beijing 100053 China;

    Department of Radiology Beijing Anzhen Hospital Capital Medical University Beijing 100029 China;

    Institute of Acupuncture and Moxibustion China Academy of Chinese Medical Sciences Beijing 100171 China;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    High pitch; Dual-source CT; Coronary angiography; Electrocardiography; Image quality;

    机译:高音高;双源CT;冠状动脉造影;心电图;图像质量;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号