首页> 外文期刊>IEEE Transactions on Instrumentation and Measurement >A laser-Doppler system for measuring distributions of blood particles in narrow flow channels
【24h】

A laser-Doppler system for measuring distributions of blood particles in narrow flow channels

机译:激光多普勒系统,用于测量狭窄流道中血液颗粒的分布

获取原文
获取原文并翻译 | 示例

摘要

A laser-Doppler velocimeter with microscopic probe volume has been developed in order to measure concentration distributions of particles in flowing suspensions containing, e.g., blood cells or microspheres. Experiments on flow through simple geometries were performed to demonstrate the applicability of the apparatus. Velocity profiles could be measured with an inaccuracy of about 2% and a spatial resolution of 15 /spl mu/m, whereas concentration measurements yielded typical inaccuracies of 7%. The determination of absolute particle concentrations without additional calibration was demonstrated to be feasible too.
机译:为了测量包含例如血细胞或微球的流动悬浮液中颗粒的浓度分布,已经开发了具有显微探针体积的激光多普勒测速仪。进行了通过简单几何形状的流动实验,以证明该设备的适用性。速度分布的测量结果不准确度约为2%,空间分辨率为15 / spl mu / m,而浓度测量得出的典型不准确度为7%。事实证明,无需额外校准即可确定绝对颗粒浓度。