Motion-Artifact Reduction in Capacitive Heart-Rate Measurements by Adaptive Filtering

机译:通过自适应滤波减少电容性心率测量的运动伪影

获取原文
获取原文并翻译 | 示例
           

摘要

Electrocardiography (ECG) measurements are essential components in clinical diagnosis and monitoring. Conventional ECG measurements produce discomfort to the patients due to the use of gel and direct skin contact. Capacitive electrodes can measure ECG signals through an isolation layer; they are especially suitable for long-term ambulatory monitoring. However, capacitive ECG measurements are severely affected by motion artifacts (MAs) due to variable coupling distance. Adaptive filtering has been widely used for MA reduction in ECG measurements. Unfortunately, a reference signal recorded by additional sensors is required for the existing adaptive-filtering methods, limiting their applicability in ambulatory settings. In this paper on capacitive ECG recordings, a novel adaptive-filtering method is proposed for MA removal, where the reference signal is extracted from the power-line interference (PLI). PLI is particularly evident in a capacitive ECG due to unbalanced coupling capacitance. Along with MAs, electrode movement causes variations in the PLI amplitude. By demodulating the PLI, a reference signal reflecting variations in the coupling capacitance can be extracted for adaptive MA removal. The proposed method was evaluated both by simulations and real data and compared with the acceleration-based adaptive-filtering method. Comparable or higher ECG signal-to-noise ratio was achieved by the proposed method with a computational cost of 79 mu s/iteration, indicating effective MA removal. The proposed method may, therefore, lead to improved analysis of capacitive ECG signals in ambulatory settings.
机译:心电图(ECG)测量是临床诊断和监测中的必要组分。由于使用凝胶和直接皮肤接触,传统的ECG测量因患者产生不适。电容电极可以通过隔离层测量ECG信号;它们特别适用于长期的动态监测。然而,由于可变耦合距离,电容式ECG测量因运动伪影(MAS)受到严重影响。自适应滤波已广泛用于ECG测量的MA降低。遗憾的是,现有的自适应过滤方法需要由附加传感器记录的参考信号,限制了它们在动态设置中的适用性。本文在电容式心电图记录上,提出了一种新的自适应滤波方法,用于MA移除,其中参考信号从电力线干扰(PLI)中提取。由于耦合电容不平衡,PLI在电容式ECG中特别明显。除了MAS,电极运动会导致PLI幅度的变化。通过解调PLI,可以提取反映耦合电容的变化的参考信号以用于自适应MA移除。通过模拟和实际数据评估所提出的方法,并与基于加速的自适应滤波方法进行比较。通过计算成本为79μS/迭代的所提出的方法,实现了可比或更高的ECG信噪比,表明有效的MA去除。因此,所提出的方法可以导致在动态设置中改善了电容式ECG信号的分析。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号