首页> 外文期刊>Instrumentation and Measurement, IEEE Transactions on >Precision AC–DC Transfer Measurement System Based on a 1000-V Inductive Voltage Divider
【24h】

Precision AC–DC Transfer Measurement System Based on a 1000-V Inductive Voltage Divider

机译:基于1000V电感分压器的精密AC-DC传输测量系统

获取原文
获取原文并翻译 | 示例

摘要

The traceability for thermal voltage converters (TVCs) usually depends on a range-to-range buildup technique that requires either very small voltage-level dependencies of the converters or precise knowledge of these level dependencies. In practice, this inevitably leads to uncertainties that accumulate in the buildup process and dominate the overall uncertainty of the high-voltage ac–dc transfer measurement. This paper presents an alternative technique that uses a 1000-V inductive voltage divider (IVD) developed at the National Measurement Institute, Linfield, Australia (NMIA). The use of the IVD eliminates the need for the buildup, which leads to a significant reduction in the uncertainty. The standard uncertainty of the new high-voltage TVC calibration setup is on the order of 1 $muhbox{V/V}$, which is significantly smaller than the uncertainties derived from the buildup technique.
机译:热电压转换器(TVC)的可追溯性通常取决于范围到范围的构建技术,该技术需要转换器的很小的电压电平依赖性或对这些电平依赖性的精确了解。在实践中,这不可避免地导致不确定性在建立过程中累积,并主导着高压交流交直流测量的总体不确定性。本文提出了一种替代技术,该技术使用了澳大利亚林菲尔德国家测量研究所(NMIA)开发的1000V电感分压器(IVD)。 IVD的使用消除了积累的需要,从而大大降低了不确定性。新的高压TVC校准设置的标准不确定度约为1 muhbox {V / V} $,大大小于从累积技术得出的不确定度。