首页> 外文期刊>IEICE Transactions on Electronics >Spectrophotometer Calibration by a Double Integrating Sphere Reference Light Source and Display Panel Measurement Using Dark Sphere
【24h】

Spectrophotometer Calibration by a Double Integrating Sphere Reference Light Source and Display Panel Measurement Using Dark Sphere

机译:通过双积分球参考光源进行分光光度计校准,并使用暗球进行显示面板测量

获取原文
获取原文并翻译 | 示例

摘要

We succeeded to develop a reference light source in the range of very low luminance using a double integrating sphere system, and calibrated a commercial spectrophotometer below 1 × 10~(-5) cd/m~2 levels, which is 1/100 lower than the specified limit for measurement. And we improved measurements in the ultra low luminance range of displays using the calibrated commercial spectrophotometer and a dark sphere to suppress the influence of the surround.
机译:我们成功地使用双积分球系统开发了一种在极低亮度范围内的参考光源,并校准了低于1×10〜(-5)cd / m〜2水平的商用分光光度计,该水平比1×10〜(-5)cd / m〜2低。指定的测量极限。并且,我们使用校准的商用分光光度计和暗球改善了显示器在超低亮度范围内的测量,以抑制周围环境的影响。