Histotripsy fractionates soft tissue by well-controlled acoustic cavitation using microsecond-long, high-intensity ultrasound pulses. The feasibility of using histotripsy as a noninvasive, drug-free, and image-guided thrombolysis method has been shown previously. A new histotripsy approach, termed microtripsy, has recently been investigated for the thrombolysis application to improve treatment accuracy and avoid potential vessel damage. In this study, we investigated the effects of pulse repetition frequency (PRF) on microtripsy thrombolysis. Microtripsy thrombolysis treatments using different PRFs (5, 50, and 100 Hz) and doses (20, 50, and 100 pulses) were performed on blood clots in an in vitro vessel flow model. To quantitatively evaluate the microtripsy thrombolysis effect, the location of focal cavitation, the incident rate of pre-focal cavitation on the vessel wall, the size and location of the resulting flow channel, and the generated clot debris particles were measured. The results demonstrated that focal cavitation was always well confined in the vessel lumen without contacting the vessel wall for all PRFs. Pre-focal cavitation on the front vessel wall was never observed at 5Hz PRF, but occasionally observed at PRFs of 50 Hz (1.2%) and 100 Hz (5.4%). However, the observed pre-focal cavitation was weak and did not significantly affect the focal cavitation. Results further demonstrated that, although the extent of clot fractionation per pulse was the highest at 5 Hz PRF at the beginning of treatment (
机译:通过使用微秒长的高强度超声脉冲,通过良好控制的声空化,组织碎裂可将软组织分级。先前已经证明了使用组织曲霉作为无创,无药物和图像引导的溶栓方法的可行性。最近,针对溶栓应用研究了一种新的称为三重曲张的组织曲张疗法,以提高治疗准确性并避免潜在的血管损伤。在这项研究中,我们调查了脉冲重复频率(PRF)对微三段血栓溶解的影响。在体外血管流动模型中,对血块进行了使用不同PRF(5、50和100 Hz)和剂量(20、50和100脉冲)的微三段溶栓治疗。为了定量评估微三段溶栓作用,局部空化的位置,聚焦前空化在血管壁上的入射率,所得流道的大小和位置以及产生的血块碎片颗粒,进行了测量。结果表明,对于所有PRF,局灶性空化始终很好地局限在血管腔中,而不接触血管壁。在5Hz PRF下从未观察到前血管壁发生焦前空化,但偶尔在50Hz(1.2%)和100Hz(5.4%)的PRF下观察到。但是,观察到的聚焦前空化作用很弱,并且没有显着影响聚焦空化作用。结果进一步表明,尽管在治疗开始时每个脉冲的凝块分级程度在5 Hz PRF时最高(

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号