This work presents a biopotential front-end amplifier in which the MOS transistors are biased in subthreshold region with a supply voltage and current of 0.4–0.8 V and 0.23–1.86 $mu$A, respectively, to reduce the system power. Flicker noise is then removed using a chopping technique, and differential interference produced by electrode impedance imbalance is suppressed using a Gm-C filter. Additionally, the circuit is fabricated using TSMC 0.18 $mu$m CMOS technology with a core area of 0.77$,times,$ 0.36 mm$^{2}$ . With a minimum supply voltage of 0.4 V, the measured SNR and power consumption of the proposed IC chip are 54.1 dB and 0.09 $mu$W , respectively.
机译:这项工作提出了一种生物电势前端放大器,其中MOS晶体管分别在亚阈值区域内偏置,其电源电压和电流分别为0.4-0.8 V和0.23-1.86μA/ A,以降低系统功率。然后使用斩波技术消除闪烁噪声,并使用Gm-C滤波器抑制由电极阻抗不平衡产生的差分干扰。此外,该电路是使用TSMC 0.18μmCMOS技术制造的,其核心面积为0.77×0.36 mm2 ^ 2。在最小电源电压为0.4 V的情况下,建议的IC芯片的实测SNR和功耗分别为54.1 dB和0.09μW。