An ECG Recording Front-End With Continuous-Time Level-Crossing Sampling

机译:具有连续时间水平交叉采样的ECG记录前端

获取原文
获取原文并翻译 | 示例

摘要

An ECG recording front-end with a continuous- time asynchronous level-crossing analog-to-digital converter (LC-ADC) is proposed. The system is a voltage and current mixed-mode system, which comprises a low noise amplifier (LNA), a programmable voltage-to-current converter (PVCC) as a programmable gain amplifier (PGA) and an LC-ADC with calibration DACs and an RC oscillator. The LNA shows an input referred noise of 3.77 over 0.06 Hz–950 Hz bandwidth. The total harmonic distortion (THD) of the LNA is 0.15% for a 10 input. The ECG front-end consumes 8.49 from a 1 V supply and achieves an ENOB up to 8 bits. The core area of the proposed front-end is , fabricated in a 0.18 CMOS technology.
机译:提出了一种具有连续时间异步电平交叉模数转换器(LC-ADC)的ECG记录前端。该系统是电压和电流混合模式系统,包括一个低噪声放大器(LNA),一个作为可编程增益放大器(PGA)的可编程电压-电流转换器(PVCC)和一个带有校准DAC的LC-ADC和RC振荡器。 LNA在0.06 Hz至950 Hz带宽上显示的输入参考噪声为3.77。对于10个输入,LNA的总谐波失真(THD)为0.15%。 ECG前端通过1 V电源消耗8.49,并实现高达8位的ENOB。拟议前端的核心区域是采用0.18 CMOS技术制造的。