首页> 外文期刊>IEEE transactions on biomedical circuits and systems >A Magnetic Plethysmograph Probe for Local Pulse Wave Velocity Measurement
【24h】

A Magnetic Plethysmograph Probe for Local Pulse Wave Velocity Measurement

机译:用于局部脉搏波速度测量的磁性脉搏波描记器探头

获取原文
获取原文并翻译 | 示例

摘要

Objective: We present the design and experimental validation of an arterial compliance probe with dual magnetic plethysmograph (MPG) transducers for local pulse wave velocity (PWV) measurement. The MPG transducers (positioned at 23 mm distance apart) utilizes Hall-effect sensors and permanent magnets for arterial blood pulse detection. Methods: The MPG probe was initially validated on an arterial flow phantom using a reference method. Further, 20 normotensive subjects (14 males, age = 24 ± 3.5 years) were studied under two different physical conditions: 1) Physically relaxed condition, 2) Postexercise condition. Local PWV was measured from the left carotid artery using the MPG probe. Brachial blood pressure (BP) was measured to investigate the correlation of BP with local PWV. Results: The proposed MPG arterial compliance probe was capable of detecting high-fidelity blood pulse waveforms. Reliable local pulse transit time estimates were assessed by the developed measurement system. Beat-by-beat local PWV was measured from multiple subjects under different physical conditions. A profound increment was observed in the carotid local PWV for all subjects after exercise (average increment = 0.42 ± 0.22 m/s). Local PWV values and brachial BP parameters were significantly correlated (r ≥ 0.72), except for pulse pressure (r = 0.42). Conclusion: MPG arterial compliance probe for local PWV measurement was validated. Carotid local PWV measurement, its variations due to physical exercise and correlation with BP levels were examined during the in vivo study. Significance: A novel dual MPG probe for local PWV measurement and potential use in cuffless BP measurement.
机译:目的:我们介绍了一种用于双脉冲体积描记器(MPG)的用于局部脉搏波速度(PWV)测量的动脉顺应性探头的设计和实验验证。 MPG传感器(相距23 mm)使用霍尔效应传感器和永磁体检测动脉血脉。方法:MPG探针最初使用参考方法在动脉流模型上进行验证。此外,在两种不同的身体状况下研究了20名血压正常的受试者(14名男性,年龄= 24±3.5岁):1)身体放松的状况,2)运动后的状况。使用MPG探针从左颈动脉测量局部PWV。测量肱动脉血压(BP),以研究BP与局部PWV的相关性。结果:提出的MPG动脉顺应性探针能够检测高保真的脉搏波形。通过开发的测量系统评估了可靠的本地脉冲传输时间估计。从多个受试者在不同的身体条件下测量逐搏局部PWV。运动后所有受试者的颈动脉局部PWV均出现了明显的增加(平均增加= 0.42±0.22 m / s)。除脉压(r = 0.42)外,局部PWV值与肱BP参数显着相关(r≥0.72)。结论:MPG动脉顺应性探针可用于局部PWV测量。在体内研究过程中检查了颈动脉局部PWV测量值,其因体育锻炼而产生的变化以及与BP水平的相关性。意义:一种新颖的双MPG探头,可用于局部PWV测量,并可用于无袖BP测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号