Early automatic remote detection of combined lead insulation defect and ICD damage

机译:早期自动远程检测组合的铅绝缘缺陷和ICD损坏

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Lead and implantable cardioverter defibrillator (ICD) device failure is a severe problem in ICD therapy and may occur without preceding signs of deterioration. Insulation lead failure and subsequent ICD defect 7 months after ICD implantation for secondary prevention of sudden cardiac death (SCD) in a 70-year-old male was automatically detected with the Home Monitoring system. Immediate lead and device replacement was performed. This case illustrates the benefit of permanent automatic remote monitoring of implanted active devices.
机译:铅和植入式心脏复律除颤器(ICD)设备故障是ICD治疗中的严重问题,可能在没有恶化迹象的情况下发生。使用家庭监测系统自动检测到ICD植入后7个月,用于二级预防心源性猝死(SCD)的7例患者的绝缘导线故障和随后的ICD缺陷。立即进行引线和设备更换。这种情况说明了对植入的有源设备进行永久自动远程监视的好处。