首页> 外文期刊>JA Clinical Reports >Spurious electroencephalographic activity due to pulsation artifact in the depth of anesthesia monitor
【24h】

Spurious electroencephalographic activity due to pulsation artifact in the depth of anesthesia monitor

机译:由于麻醉显示器深度脉动伪影引起的杂散脑电图活动

获取原文
           

摘要

BackgroundThe depth of anesthesia (DOA) is estimated based on the anesthesia-induced electroencephalogram (EEG) changes. However, the surgical environment, as well as the patient him/herself, generates electrical interferences that cause EEG waveform distortion.Case presentationA 52-year-old patient required general anesthesia due to the right femur necrotizing fasciitis. He had no history of epilepsy or head injury. His cardiovascular status was stable without arrhythmia under propofol and remifentanil anesthesia. The DOA was evaluated with Root? with SedLine? Brain Function Monitoring (Masimo Inc, Irvine, CA). The EEG showed a rhythmic, heart rate time-locked pulsation artifact, which diminished after electrode repositioning. Offline analysis revealed that the pulse wave-like interference in EEG was observed at the heart rate frequency.ConclusionsWe experienced an anesthesia case that involves a pulsation artifact generated by the superficial temporal artery contaminating the EEG signal. Numerous clinical conditions, including pulsation artifact, disturb anesthesia EEG.
机译:背景技术基于麻醉诱导的脑电图(EEG)变化估计麻醉深度(DOA)。然而,外科环境以及患者他/她自己,产生导致EEG波形失真的电气干扰。CASE呈现52岁的患者由于正确的股骨坏死性筋膜炎,需要全身麻醉。他没有癫痫或头部受伤的历史。他的心血管状态在异丙酚和雷芬丹尼尔麻醉下没有心律失常的稳定性。 DOA被root评估?用塞线?脑功能监测(Masimo Inc,Irvine,CA)。脑电图显示了电极重新定位后减少的节奏,心率时锁的脉动伪影。离线分析显示,在心率频率下观察到脑电图的脉搏波状干扰。结论我们经历了一种麻醉案例,涉及污染脑电图的浅表颞动脉产生的脉动伪像。许多临床条件,包括脉动伪影,干扰麻醉脑电图。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号