Spurious electroencephalographic activity due to pulsation artifact in the depth of anesthesia monitor

机译:由于麻醉监测深度的脉动伪影引起的杂散脑电图活动

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Normalization of electroencephalography after L2 electrode repositioning. a A snapshot of the Root® with SedLine® Brain Function Monitoring (Masimo Inc, Irvine, CA) is presented. A rhythmic, heart rate time-locked pulsation artifact is overlying L2. b Following L2 electrode repositioning, a pulsation artifact has completely disappeared. All four electroencephalogram (EEG) traces show a burst-suppression pattern. c Two 10-s samples (left and right) of four EEG channels (L1, R1, L2, and R2), electrocardiogram (ECG), and plethysmography (pleth) show how the waveform of L2 changes after electrode replacement. The synchronization between ECG/pleth versus EEG is not perfect due to separate standalone monitoring devices, but it can be seen that the pace of the low fluctuation in L2 EEG is similar to that of the heart rate and pleth. The units of EEG and ECG signals are microvolts. d Heart rate (red) and instantaneous frequency of the low-frequency component of L2. Both sample-by-sample estimate and smoothed version over a 4-s window are shown in black color. The time axis values correspond to those in c above. The Roman numerals I and II indicate the timepoints of making the start of the electrode replacement (I) and recovery of the EEG signal (II), respectively

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号