A comparison of the analysis of 3 types of body fluids using the XN-350 hematology analyzer versus light microscopy assessment

机译:使用XN-350血液分析仪对3种体液分析的比较与光学显微镜评估

获取原文
获取外文期刊封面目录资料

摘要

ABSTRACT:We evaluated the capacity of the XN-350 instrument to analyze 3 different types of body fluid samples under "body fluid mode."The performance of XN-350 was evaluated in terms of precision, carryover, limit of blank, limit of detection, limit of quantification, and linearity. Cell enumeration and differential data produced by the XN-350 were compared to manual chamber counting results in 63 cerebrospinal fluid (CSF), 51 ascitic fluid, and 51 pleural fluid (PF) samples. Comparisons between XN-350 versus Cytospin data were also performed in PF samples.The precision, carry-over, limit of blank, and linearity of the XN-350 were acceptable. The limits of detection for white blood cells (WBCs) and red blood cells were 1.0/μL, and 1,000.0/μL, respectively; the corresponding limits of quantitation (LOQs) were 5.0/μL and 2,000.0/μL, respectively. The XN-350's cell enumeration and differential counting correlated well with those of manual chamber counting for all 3 sample types (except for differential counting in CSF samples), particularly parameters involving monocytes (r?=?0.33) and mononuclear cells (MO- body fluid [BF]; r?=?0.26), as well as total cell (TC-BF) enumeration (r?=?0.50) and WBC-BF (r?=?0.50) in PF samples. The MO-BF in CSF samples differed significantly from manual chamber counting results, but neither TC-BF nor WBC-BF in PF samples did. The XN-350 also showed good correlations with Cytospin analyses for differential counting of neutrophils, lymphocytes, and monocytes in PF samples. The differential counting of eosinophils via the XN-350 and Cytospin were not significantly correlated, but the difference between them was not significant.The XN-350 is an acceptable alternative to manual fluid analysis. Samples with low cellularity around the LOQ should be checked manually. Moreover, manual differential counting should be performed on CSF samples, particularity those with low cell numbers.Copyright ? 2021 the Author(s). Published by Wolters Kluwer Health, Inc.
机译:摘要:我们评估了XN-350仪器的能力,分析了“体液模式”下的3种不同类型的体液样品。“XN-350的性能在精度,携带,空白极限,检测限的限度方面进行了评估,量化限制,线性。将XN-350产生的细胞枚举和差分数据进行比较,以便手动腔室计数导致63脑脊液(CSF),51个腹水和51个胸膜液(PF)样品进行比较。在PF样品中也进行了XN-350与细胞螺旋素数据之间的比较。XN-350的精度,随机,空白极限和线性度是可接受的。白细胞(WBCS)和红细胞检测的极限分别为1.0 /μl,和1,000.0 /μl;相应的定量限制(LOQ)分别为5.0 /μl和2,000.0 /μl。 XN-350的细胞枚举和差分计数与所有3种样品类型的手动腔室的计数良好(除了CSF样本中的差分计数除外),特别是涉及单核细胞的参数(R?= 0.33)和单核细胞(Mo-体流体[bf]; r?= 0.26),以及PF样品中的总细胞(TC-BF)枚举(R?=Δ0.50)和WBC-BF(R?= 0.50)。 CSF样品中的MO-BF从手动腔室计数结果中显着不同,但PF样本中的TC-BF和WBC-BF都不存在。 XN-350还表现出与PF样品中的中性粒细胞,淋巴细胞和单核细胞的差异计数的细胞螺旋素分析良好的相关性。通过XN-350和Cytospin的嗜酸性粒细胞的差异计数没有显着相关,但它们之间的差异不是显着的。XN-350是手动流体分析的可接受的替代方案。应手动检查LOQ周围低细胞性的样品。此外,应对CSF样本进行手动差分计数,特别是具有低单元号的特殊性.Copyright? 2021提交人。由Wolters Kluwer Health,Inc。出版

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号