Quantitative Measurement of Plasma Free Metanephrines by a Simple and Cost-Effective Microextraction Packed Sorbent with Porous Graphitic Carbon and Liquid Chromatography-Tandem Mass Spectrometry

机译:用多孔石墨碳和液相色谱 - 串联质谱法进行简单且经济高效的微萃取的血浆可自由萃取剂的定量测量。

获取原文
           

摘要

Plasma free metanephrines are widely used for the diagnosis of pheochromocytoma and paraganglioma (PPGL), yet quantifying metanephrines using a simple and cost-effective approach may be challenging due to preanalytical and analytical constraints. In this study, we established and validated a new method for quantitative measurement of plasma free metanephrines based on microextraction by packed sorbent (MEPS) with porous graphitic carbon (PGC) and liquid chromatography-tandem mass spectrometry (HILIC-MS/MS). The elution step was fully compatible with HILIC mode without evaporation and reconstitution. The analytes were well resolved, and potential interferences (54 substances) were investigated. This method was linear from 24.7–2717?pg/mL for metanephrine (MN) and 24.5–4010?pg/mL for normetanephrine (NMN) with a coefficient of determination ( R 2 ) higher than 0.994. The limit of MN and NMN detection were 12.4?pg/mL and 12.3?pg/mL, respectively. The intra- and interassay impressions were ≤12.8% for spiked quality controls and ≤13.6% for commercial quality controls; the method recoveries ranged within 88.0–109.0%, respectively. The area under the receiver operating characteristic (ROC) curve was 0.848?±?0.047 for MN and 0.979?±?0.021 for NMN. Validation that was performed by comparing clinical specimens with various biochemical results showed that plasma free metanephrines in a seated position had comparable sensitivity and lower specificity to urinary free metanephrines, which could be compensated by combining other biochemical tests. The newly developed MEPS method resulted as a time-saving, reliable, and cost-effective microextraction technique that can be applied for a successful screening of PPGL.
机译:血浆自由Metanephrines广泛用于诊断嗜铬细胞瘤和Papanglioma(PPGL),但使用简单且经济有效的方法量化Metanephrines可能是由于预期和分析约束而挑战。在这项研究中,我们建立并验证了一种基于多孔石墨(PGC)和液相色谱 - 串联质谱(HILIC-MS / MS)的填充吸附剂(MEP)基于微萃取的微萃取的血浆自由metanephrines的量度测量新方法。洗脱步骤与HILIC模式完全兼容而不蒸发和重建。分析物溶解良好,研究了潜在的干扰(54种物质)。该方法从24.7-2717〜27-2717〜27-2717〜24.5-4010〜24.5-4010〜40〜4010·pg / ml的常规胰岛(NMN)线性的线性是线性的,具有高于0.994的测定系数(R 2)。 Mn和NMN检测的极限分别为12.4〜pg / ml和12.3〜pg / ml。对于商业质量控制,尖刺质量控制的内部和中间印象≤12.8%,≤13.6%;该方法回收率分别在88.0-109.0%内。接收器操作特性(ROC)曲线下的区域为0.848°α±0.047,适用于NMN的0.979±0.021±0.021。通过比较具有各种生化结果的临床标本进行的验证表明,坐姿的血浆可自由metanephers对尿液定位的敏感性和更低的特异性,可以通过组合其他生化测试来补偿。新开发的MEPS方法导致了一种节省的可靠性和经济高效的微萃取技术,可以应用于成功筛选PPGL。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号