Objectives To compare the performance of two syndromic panels: Luminex xTAG Gastrointestinal Pathogen Panel (GPP) and FilmArray Gastrointestinal (GI) panel. Methods A total of 243 diarrhea specimens were detected by two panels in parallel, and the inconsistent results were analyzed by real-time PCR or reverse transcription PCR (RT-PCR). The target concentration in specimens was examined by comparing the crossing point values of FilmArray, the median fluorescence intensity of xTAG and the cycle threshold values in any discrepancies. Results For pathogens detected by both panels, the positive rates of FilmArray GI and xTAG GPP were 65.0% and 48.6%, respectively. The two panels showed high consistency (kappa ≥0.74) in detecting norovirus, rotavirus and Campylobacter , while there was low consistency (kappa ≤0.40) in detecting Cryptosporidium , Salmonella , Shiga toxin-producing Escherichia coli (STEC) and enterotoxigenic Escherichia coli (ETEC). Samples with low concentration targets were more often detected by FilmArray than with xTAG GPP. The xTAG GPP was more likely to be affected by amplification inhibitors. Several defects of xTAG GPP were found in detecting ETEC. Conclusions FilmArray was more sensitive. For specimens with low target concentrations or containing ETEC heat stable enterotoxin, the false negatives of xTAG GPP need to be considered.
机译:目标可以比较两种综合组面板的性能:Luminex XTAG胃肠道病原体面板(GPP)和Fimbarray胃肠道(GI)面板。方法通过两个面板并行检测总共243个腹泻样品,通过实时PCR或逆转录PCR(RT-PCR)分析不一致的结果。通过比较FetParray的交叉点值,XTAG的中值荧光强度和任何差异的循环阈值来检查标本中的目标浓度。由两个面板检测到病原体的结果,薄膜阵列GI和XTAG GPP的阳性率分别为65.0%和48.6%。这两个面板在检测诺罗病毒,轮状病病毒和弯曲杆菌中显示出高一致性(Kappa≥0.74),而在检测Cryptosporidium,Salmonella,Shiga毒素的大肠杆菌(STEC)和肠毒素大肠杆菌(ETEC)中存在低一致性(Kappa≤0.40) )。薄膜阵列比用XTAG GPP更常用具有低浓度靶标的样品。 XTAG GPP更可能受扩增抑制剂的影响。发现XTAG GPP的几种缺陷在检测etec中被发现。结论薄膜阵列更敏感。对于靶浓度低或含有ETEC热稳定肠毒素的标本,需要考虑XTAG GPP的假阴性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号