Validation of novel automatic ultra-wideband radar for sleep apnea detection

机译:验证睡眠呼吸暂停检测的新型自动超宽带雷达

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Background: To validate the accuracy of ultra-wideband (UWB) wireless radar for the screening diagnosis of sleep apnea. Methods: One hundred and seventy-six qualified participants were successfully recruited. Apnea-hypopnea index (AHI) results from polysomnography (PSG) were reviewed by physicians, while the radar device automatically calculated AHI values with an embedded chip. All results were statistically analyzed. Results: A UWB radar-based AHI algorithm was successfully developed according to respiratory movement and body motion signals. Of all 176 participants, 63 exhibited normal results (AHI 5/hr) and the remaining 113 were diagnosed with obstructive sleep apnea. Significant correlation was detected between radar AHI and PSG AHI (Intraclass correlation coefficient 0.98, P0.001). Receiver operating characteristic curve (ROC) analysis revealed high sensitivity and specificity. High concordance in participants with varying gender, age, BMI, and PSG AHI was reached. Conclusions: The UWB radar may be a portable, convenient, and reliable device for obstructive sleep apnea screening.
机译:背景:验证超宽带(UWB)无线雷达的准确性,以便筛选睡眠呼吸暂停的诊断。方法:成功招募了一百七十六合格的参与者。 PolySomNography(PSG)的呼吸暂停缺氧索引(AHI)由医生审查,而雷达设备自动计算嵌入式芯片的AHI值。所有结果都在统计分析。结果:根据呼吸运动和车身运动信号成功开发了基于UWB雷达的AHI算法。在所有176名参与者中,63种表现出正常结果(AHI <5 / HR),剩余113次被诊断出阻塞性睡眠呼吸暂停。在雷达AHI和PSG AHI之间检测到显着的相关性(颅内相关系数0.98,P <0.001)。接收器操作特征曲线(ROC)分析显示出高敏感性和特异性。达到了不同性别,年龄,BMI和PSG AHI的参与者的高度一致。结论:UWB雷达可以是便携式,方便,可靠的装置,用于阻塞性睡眠呼吸暂停筛选。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号