Clinical Evaluation of the Cell-Dyn? 1700CS Blood Counter,

机译:Cell-Dyn?的临床评估1700CS血液计数器

获取原文
           

摘要

New generations of hematology analyzers produce fast and reliable data on blood count, while at the same time offering screening information on differential blood count. The aim of this study was to evaluate a Cell-Dyn? (CD) 1700CS (Abbott Diagnostics, Abbott Park, IL) hematology analyzer that determines 18 parameters, including three-part differential blood count.A single unit instrument consists of a sample analyzer, data module section, and printer as a separate component. The analyzer contains the hardware to aspirate, dilute, and analyze each whole-blood sample from open and closed collection tubes or prediluted samples. Aspiration volume in open mode is 30 μL, in closed mode 450 μL, and in predilute mode 40 μL. The data module section includes a computer controlling all operations and storing a total of 5000 numeric and graphic data results, color video display monitor, and keyboard. The keyboard allows entry of sample identification number, patient name, sex, date of birth, physician, collection date and time, and comments. In addition, limits and reference ranges can be printed with each patient’s results.Counting of blood cells is based on the volumetric impedance method, directly measuring white blood cells (WBC), red blood cells (RBC), hemoglobin (HGB), platelets, mean corpuscular volume (MCV), and mean platelet volume, and automatically calculating hematocrit (HCT), mean corpuscular hemoglobin (MCH), MCH concentration (MCHC), RBC distribution width, plateletcrit, and platelet distribution width. The instrument differentiates the subpopulations of lymphocytes, granulocytes, and the mid-cell fraction (eosinophils, basophils, monocytes, and precursors of WBCs) by electronic sizing. Specially formulated reagents cause the WBC membrane to shrink around the nucleus while keeping the cell intact, allowing separation of white cells according to their volume. Lymphocytes fall within the small-cell region, neutrophils within the large-cell region, and the remaining cells into the mid-size cell region. The three-part differential screen is provided …
机译:新一代血液学分析仪可生成快速可靠的血细胞计数数据,同时提供筛查差异血细胞计数信息。这项研究的目的是评估Cell-Dyn? (CD)1700CS(伊利诺伊州阿伯特帕克市的Abbott Diagnostics)血液分析仪可确定18个参数,包括三部分差分血细胞计数。单个仪器包括一个样品分析仪,数据模块部分和打印机作为单独的组件。分析仪包含用于抽吸,稀释和分析来自开放式和封闭式收集管或预稀释样品中的每个全血样品的硬件。在打开模式下的抽吸量为30μL,在关闭模式下的抽吸量为450μL,在预稀释模式下的抽吸量为40μL。数据模块部分包括一台控制所有操作并存储总共5000个数字和图形数据结果的计算机,彩色视频显示监视器和键盘。键盘允许输入样品识别号,患者姓名,性别,出生日期,医师,采集日期和时间以及注释。此外,每个患者的结果都可以打印出极限值和参考范围。血细胞计数基于体积阻抗法,直接测量白细胞(WBC),红细胞(RBC),血红蛋白(HGB),血小板,平均红细胞体积(MCV)和平均血小板体积,并自动计算血细胞比容(HCT),平均红细胞血红蛋白(MCH),MCH浓度(MCHC),RBC分布宽度,血小板压积和血小板分布宽度。该仪器通过电子上浆区分淋巴细胞,粒细胞和中层细胞部分(嗜酸性粒细胞,嗜碱性粒细胞,单核细胞和白细胞的前体)的亚群。特殊配制的试剂可使WBC膜在细胞核周围收缩,同时保持细胞完整,从而根据其体积分离白细胞。淋巴细胞落在小细胞区域内,嗜中性粒细胞在大细胞区域内,其余细胞进入中型细胞区域。提供三部分的差分屏幕…

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号