Two bone alkaline phosphatase assays compared with osteocalcin as a marker of bone formation in healthy elderly women.

机译:将两种骨碱性磷酸酶测定法与骨钙素比较,作为健康老年妇女的骨形成指标。

获取原文
           

摘要

Serum bone alkaline phosphatase (ALP; EC 3.1.3.1) was measured with a wheat germ agglutinin (WGA) precipitation assay and with a new IRMA in a group of healthy elderly women. Both assays were correlated with serum total ALP activity and with osteocalcin. The two bone ALP assays have comparable within- and between-run imprecisions (WGA assay within-run CVs 2.6-5.4% and between-run, 4.0-5.1%; IRMA within-run CV 5.0% and between-run, 3.2%). Comparison of the WGA precipitation assay (x) with the IRMA (y) demonstrated a correlation coefficient of 0.87 [Deming regression equation: y = (0.58 +/- 0.02)x - (4.62 +/- 0.45); n = 101; Sy/x = 1.26; P 0.001). Correlation studies with osteocalcin and total ALP showed correlation coefficients (all P 0.001) of 0.34 and 0.65, respectively, for the WGA precipitation assay and of 0.36 and 0.68, respectively, for the IRMA. We conclude that the two bone ALP assays have similar imprecision and that neither can be given preference over the other as a marker of bone turnover.
机译:在一组健康的老年妇女中,通过小麦胚芽凝集素(WGA)沉淀测定法和新型IRMA测定了血清骨碱性磷酸酶(ALP; EC 3.1.3.1)。两种测定均与血清总ALP活性和骨钙素相关。两种骨ALP检测在运行中和运行中的不精确度均相当(WGA检测在运行中CV的为2.6-5.4%,在运行中为4.0-5.1%; IRMA在运行中的CV为5.0%,在运行中为3.2%) 。 WGA沉淀分析(x)与IRMA(y)的比较表明相关系数为0.87 [戴明回归方程:y =(0.58 +/- 0.02)x-(4.62 +/- 0.45); n = 101; Sy / x = 1.26; P <0.001)。与骨钙素和总ALP的相关性研究显示,WGA沉淀测定的相关系数(所有P <0.001)分别为0.34和0.65,IRMA的相关系数分别为0.36和0.68。我们得出的结论是,两种骨ALP测定法具有相似的不精确性,并且不能将两者作为骨转换的标志物而给予优先。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号