首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Direct Potentiometric Determination of Potassium and Sodium in Blood, Plasma, and Serum
【24h】

Direct Potentiometric Determination of Potassium and Sodium in Blood, Plasma, and Serum

机译:直接电位法测定血液,血浆和血清中的钾和钠

获取原文
获取外文期刊封面目录资料

摘要

A potentiometric analyzer, developed to directly measure potassium and sodium concentrations, was evaluated on aqueous solutions, whole blood, serum, and plasma. The analyzer consists of three electrodes—sodium, potassium, and reference—mounted in a flow-cell configuration, and a modified digital pH meter, which displays the ion concentrations directly. Volume of the cell is about 100 μl. The electrodes equilibrate to each new sample in less than a minute. Calibrated with aqueous standard solution, the system yields essentially the same values for whole blood and the corresponding plasma, and results agree with those obtained for the same samples of plasma by flame photometry.
机译:对水溶液,全血,血清和血浆进行了电位计分析仪的开发,可直接测量钾和钠的浓度。该分析仪由安装在流通池结构中的三个电极(钠,钾和参比电极)和改进的数字pH计组成,可直接显示离子浓度。细胞的体积为约100μl。电极在不到一分钟的时间内平衡到每个新样品。用标准水溶液校准后,该系统对全血和相应的血浆产生基本相同的值,结果与通过火焰光度法对相同血浆样品获得的结果一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号