首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Determination of serum aspartate aminotransferase with pyridoxal 5'-phosphate in the Technicon SMAC and Du Pont aca compared and correlated with the IFCC recommended method.
【24h】

Determination of serum aspartate aminotransferase with pyridoxal 5'-phosphate in the Technicon SMAC and Du Pont aca compared and correlated with the IFCC recommended method.

机译:与Technicon SMAC和Du Pont aca中的吡ido醛5'-磷酸测定血清天冬氨酸转氨酶的比较,并与IFCC推荐的方法相关联。

获取原文

摘要

Inexpensive improvements in continuous-flow analytical apparatus can eliminate some perplexing inconsistencies and limitations in the use of the Technicon SMAC instrument for measuring the activity in serum of aspartate aminotransferase (EC 2.6.1.1) in the presence of pyridoxal 5'-phosphate. In addition, a calibration procedure based on values for this enzyme obtained by a modified IFCC method can be used to calibrate both the Technicon SMAC and the Du Pont aca instruments to produce excellent correlation between the two.
机译:连续流分析仪的廉价改进可以消除使用Technicon SMAC仪器在5'-磷酸吡ido醛存在下测量天冬氨酸转氨酶(EC 2.6.1.1)血清活性时的一些困惑和局限性。另外,基于通过改良的IFCC方法获得的该酶的值的校准程序可用于校准Technicon SMAC和Du Pont aca仪器,从而在两者之间产生出色的相关性。