The multilayered film analyzer: glucose in serum, plasma, cerebrospinal fluid, and urine; and urea nitrogen in serum and plasma.

机译:多层膜分析仪:血清,血浆,脑脊液和尿液中的葡萄糖;以及血清和血浆中的尿素氮。

获取原文

摘要

Some basic laboratory performance criteria were studied for the Eastman Kodak glucose and urea nitrogen analyzer. Serum, plasma, urine, and cerebrospinal fluid were tested. Precision, both "within-day" and "between-day," for both analytes was less than 2.2% (CV). Both analytes demonstrated linearity, with about 100% recovery of added substrates. Comparisons with continuous-flow procedures demonstrated good correlation. A variety of sera and plasmas can be used for glucose analysis; oxalate--fluoride-treated plasma is unacceptable for use in analysis for urea nitrogen. We saw no effects on glucose or urea nitrogen analysis from hemolysis, lipemia, icterus, some common drugs, ammonia, or abnormally high protein concentrations. Minimum amounts detectable were: glucose (serum) 104 mg/L; glucose (cerebrospinal fluid) 100 mg/L; and urea nitrogen (serum) 21 mg/L. Calibration procedures are discussed for protein-based and aqueous samples.
机译:研究了伊士曼柯达葡萄糖和尿素氮分析仪的一些基本实验室性能标准。测试了血清,血浆,尿液和脑脊液。两种分析物的“日内”和“日间”精度均低于2.2%(CV)。两种分析物均显示线性,添加的底物回收率约为100%。与连续流程序的比较显示出良好的相关性。各种血清和血浆都可以用于葡萄糖分析。草酸盐-氟化物处理的血浆不能用于尿素氮的分析。我们发现溶血,血脂,黄疸,某些常用药物,氨水或异常高的蛋白质浓度对葡萄糖或尿素氮分析没有影响。可检测的最小量为:葡萄糖(血清)104 mg / L;葡萄糖(脑脊液)100 mg / L;尿素氮(血清)为21 mg / L。讨论了基于蛋白质和水性样品的校准程序。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号