首页> 外文期刊>Bulletin of the Korean Chemical Society >Coherence Gated Three-dimensional Imaging System using Organic Photorefractive Holography
【24h】

Coherence Gated Three-dimensional Imaging System using Organic Photorefractive Holography

机译:使用有机光折变全息术的相干门控三维成像系统

获取原文
           

摘要

This paper discusses a coherence-gated three-dimensional imaging system based on photorefractive holography, which was applied to imaging through turbid media with a view to developing biomedical instrumentation. A rapid response photorefractive device doped with 2,4,7-trinitro-9-fluorenylidene malononitrile was used to generate the hologram grating. The estimated depth resolution was 20 μm, which corresponds to the coherence length of the light source. In this coherence imaging system, tomographic imaging of a 3-dimensional object composed of a 50 μm thick cylindrical layer was achieved. The proposed coherence imaging system using an organic photorefractive material can be used as an optical tomography system for biological applications
机译:本文讨论了一种基于光折射全息技术的相干门控三维成像系统,该系统被用于通过混浊介质成像,以开发生物医学仪器。掺有2,4,7-三硝基-9-芴基丙二腈的快速反应光折变器件用于产生全息光栅。估计的深度分辨率为20μm,对应于光源的相干长度。在该相干成像系统中,实现了由50μm厚的圆柱层组成的3维物体的层析成像。所提出的使用有机光折变材料的相干成像系统可以用作生物应用的光学层析成像系统