首页> 外文期刊>Bulletin of the Korean Chemical Society >Indirect Determination of Nitrite by Flame Atomic Absorption Spectrometry Using a Lead(IV) Dioxide Oxidant Microcolumn
【24h】

Indirect Determination of Nitrite by Flame Atomic Absorption Spectrometry Using a Lead(IV) Dioxide Oxidant Microcolumn

机译:二氧化铅(IV)火焰原子吸收光谱法间接测定亚硝酸盐

获取原文
           

摘要

A new, simple and fast flow injection analysis (FIA) method has been developed for the indirect determination of nitrite. The proposed indirect automatic method is based on the oxidation of nitrite to nitrate using a lead(IV) dioxide oxidant microcolumn where the flow of the sample through the microcolumn reduces the PbO2 solid phase reagent to Pb(II), which is measured by flame atomic absorption spectrometry. The absorbance of Pb(II) are proportional to the concentration of nitrite in the samples. The calibration curve was linear up to 30 mg L-1, with a detection limit of 0.11 mg L-1 for a 400 レL injected sample volume and a sampling rate of about 80 h-1. The results exhibit no interference from the presence of large amounts of ions. The developed procedure was found to be suitable for the determination of nitrite in foodstuffs and wastewaters. A relative standard deviation better than 0.9% was obtained in a repeatability study. The reliability of the method was established by parallel determination against the standard method.
机译:已经开发出一种新的,简单而快速的流动注射分析(FIA)方法来间接测定亚硝酸盐。拟议的间接自动方法是基于使用二氧化铅(IV)氧化剂微柱将亚硝酸盐氧化为硝酸盐的方法,其中样品通过微柱流将PbO 2 固相试剂还原为Pb(II) ),通过火焰原子吸收光谱法测量。 Pb(II)的吸光度与样品中亚硝酸盐的浓度成正比。最高至30 mg L -1 时,校准曲线是线性的,进样量为400レL,进样速率为0.11 mg L -1 。大约80小时 -1 。结果显示没有受到大量离子存在的干扰。发现开发的方法适用于食品和废水中亚硝酸盐的测定。在重复性研究中获得了优于0.9%的相对标准偏差。通过与标准方法平行测定来确定方法的可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号