Dosimetric verification of source strength for HDR afterloading units with Ir-192 and Co-60 photon sources: Comparison of three different international protocols

机译:用Ir-192和Co-60光子源对HDR后装单元的源强度进行剂量学验证:三种不同国际协议的比较

获取原文
           

摘要

Before clinical use of a brachytherapy source, regulations or recommendations by medical physics societies require an independent measurement of its air kerma strength by a qualified medical physicist. Currently, in addition to Ir-192, also HDR-Co-60 sources are increasingly coming into operation. However, the existing dosimetry protocols do not provide any guidelines for Co-60 sources. The purpose of this work was therefore to compare air kerma rate measurements as recommended by different dosimetry protocols for Ir-192 HDR sources and to test their applicability to Co-60 sources. Dosimetric verification of HDR afterloading source specification was performed according to three protocols, DIN 6809-2 (1993) in combination with DGMP-Report 13 (2006), IAEA-TECDOC-1274 (2002) and AAPM Report 41 (1993) for the nuclides Ir-192 and Co-60. Measurements of the sources reference air kerma rate were performed with 3 different methods (with a cylindrical chamber both in a solid phantom and in free air, and with a well chamber) and evaluated using all three protocols for each type of source and method of measurement. The measurements with all protocols and methods show deviations from the certified specification smaller than about 1.2% for Ir-192 and 2.5% for Co-60-Sources. The measurements with the well chamber showed the lowest deviations from the certificate value. Air kerma rate measurements for Co-60 HDR sources using the existing protocols are possible with accuracy sufficient to verify source calibration as provided by the source certificate. However, extension of the protocols by correction factors for measurement with Co-60 sources would be helpful.Keywords: Air kerma rate measurement for Ir-192, Co-60, brachytherapy, dosimetry
机译:在临床上使用近距离放射治疗源之前,医学物理学会的法规或建议要求合格的医学物理学家对其空气比释动能强度进行独立测量。当前,除了Ir-192外,HDR-Co-60光源也越来越多地投入使用。但是,现有的剂量学协议没有为Co-60来源提供任何指南。因此,这项工作的目的是比较不同剂量协议对Ir-192 HDR放射源推荐的空气比释动能率测量值,并测试其对Co-60放射源的适用性。根据DIN 6809-2(1993)与DGMP-Report 13(2006),IAEA-TECDOC-1274(2002)和AAPM Report 41(1993)三种核素协议对HDR后装源规格进行剂量学验证Ir-192和Co-60。使用3种不同的方法(使用实心幻影和自由空气中的圆柱室,以及使用井室)对源参考空气比释动能率进行测量,并针对所有类型的源和测量方法使用所有三种协议进行评估。所有协议和方法的测量结果表明,与Ir-192的认证规格偏差均小于1.2%,而对于Co-60-Sources的检测偏差仅为2.5%。井室的测量结果显示与证书值的偏差最小。使用现有协议对Co-60 HDR源进行空气比释动能速率测量的准确性足以验证源证书提供的源校准。然而,通过校正因子扩展协议以用于Co-60来源的测量将是有帮助的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号