Estimation of Effective Tissue Thickness in the Assessment of Plutonium-239 in Lung

机译:肺-239评估中有效组织厚度的估计

获取原文
           

摘要

For the determination of 239Pu deposited in the lungs by external counting, the absorption of low energy photons emitted from 239Pu in the chest must be taken into consideration because of its high attenuation.It is shown that the effective tissue thickness of the chest wall can be estimated by comparing the Am/Pu counting ratio-which is defined as the ratio of net counting rates between those in the Am-channel (57~63 keV) and the Pu-channel (15~19 keV)-obtained from an exposed subject with that obtained beforehand as a function of the absorber thickness of a simple chest model. An empirical expression was also derived from experiments on a phantom to estimate the effective tissue thickness. These two methods seem reasonably accurate judging from a comparison with the experimental results obtained by Rundo et al.
机译:通过外部计数确定沉积在肺中的239Pu,必须考虑从239Pu发射的低能光子在胸腔中的吸收,因为它具有很高的衰减度,这表明可以将胸壁的有效组织厚度通过比较Am / Pu计数比估算得出,Am / Pu计数比定义为从暴露对象获得的Am通道(57〜63 keV)和Pu通道(15〜19 keV)的净计数率之比。预先获得的值是简单胸部模型的吸收体厚度的函数。还从幻像上的实验中得出了经验表达式,以估计有效的组织厚度。从与Rundo等人获得的实验结果进行比较来看,这两种方法似乎相当准确。