Retrospective evaluation of exposure index (EI) values from plain radiographs reveals important considerations for quality improvement

机译:回顾性X射线照片的暴露指数(EI)值揭示了质量改进的重要考虑因素

获取原文
           

摘要

AbstractIntroductionFollowing X-ray exposure, radiographers receive immediate feedback on detector exposure in the form of the exposure index (EI).PurposeTo identify whether radiographers are meeting manufacturer-recommended EI (MREI) ranges for routine chest, abdomen and pelvis X-ray examinations under a variety of conditions and to examine factors affecting the EI.MethodsData on 5000 adult X-ray examinations including the following variables were collected: examination parameters, EI values, patient gender, date of birth, date and time of examination, grid usage and the presence of implant or prosthesis. Descriptive statistics were used to summarize each data set and the Mann–Whitney U test was used to determine significant differences, with P  0.05 indicating significance for all tests.ResultsMost examinations demonstrated EI values that were outside the MREI ranges, with significantly higher median EI values recorded for female patient radiographs than those for male patients for all manufacturers, indicating higher detector exposures for all units except for Philips digital radiography (DR), where increased EI values indicate lower exposure (P = 0.01). Median EI values for out of hours radiography were also significantly higher compared with normal working hours for all technologies (P ≤ 0.02). Significantly higher median EI values were demonstrated for Philips DR chest X-rays without as compared to those with the employment of a grid (P = 0.03), while significantly lower median EI values were recorded for Carestream Health computed radiography (CR) chest X-rays when an implant or prosthesis was present (P = 0.02).ConclusionsNon-adherence to MREIs has been demonstrated with EI value discrepancies being dependent on patient gender, time/day of exposure, grid usage and the presence of an implant or prosthesis. Retrospective evaluation of EI databases is a valuable tool to assess the need of quality improvement in routine DR.
机译:摘要简介在X射线曝光后,放射线照相技师会以曝光指数(EI)的形式立即收到检测器曝光的反馈信息。方法收集了5000次成人X射线检查的方法数据,包括以下变量:检查参数,EI值,患者性别,出生日期,检查日期和时间,网格使用情况以及检查方法。植入物或假体的存在。描述性统计数据用于汇总每个数据集,Mann-Whitney U检验用于确定显着差异,P <0.05表示所有检验均具有显着性。结果大多数检查表明EI值不在MREI范围内,中位数EI明显更高所有制造商的女性患者X射线照片记录的值均高于男性患者,这表明除Philips数字射线照相(DR)以外,所有装置的探测器曝光量均较高,其中EI值升高表示曝光量较低(P = 0.01)。与所有技术的正常工作时间相比,非工作时间放射成像的中值EI值也显着更高(P≤0.02)。与不使用网格的情况相比,未使用Philips的DR胸部X射线显示的中值EI值显着更高(P = 0.03),而Carestream Health计算X线摄影(CR)的胸部X-射线的EI中值显着降低存在植入物或假体时会发出光线(P = 0.02)结论已证明未坚持MREI,EI值差异取决于患者的性别,暴露时间/天数,使用的网格以及植入物或假体的存在。 EI数据库的回顾性评估是评估常规DR中质量改进需求的宝贵工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号