Background To compare the Mindray BC‐5180 and Sysmex XN‐1000 instruments by analyzing the results of complete blood count in the external quality assessment in Shandong Province in 2018. Methods In the external quality assessment, 10 batches of quality control materials were issued throughout the year. The test items were WBC, RBC, Hb, PLT, and HCT. The laboratories using Mindray BC‐5180 and Sysmex XN‐1000 were screened, and the results were analyzed by t test, Passing‐Bablok regression analysis, and Bland‐Altman analysis. Results Thirty‐six laboratories using Mindray BC‐5180 instruments and thirty‐six laboratories using Sysmex XN‐1000 instruments were screened, and the average difference between the two instruments results is not significant ( P ?0.05, t test). Passing‐Bablok regression analysis showed that the 95% confidence interval of the regression equation interception of each test item included 0, and the 95% confidence interval of the slope contained 1, r ?0.98, which showed that the correlation is good. The Bland‐Altman analysis showed that both instruments had more than 95% of the points within the 95% consistency limit (WBC97.2%, RBC95.6%, PLT97.2%, Hb96.7%, HCT97.5%). Within the consistency limit, the absolute value of the difference between the Mindray BC‐5180 instrument and the Sysmex XN‐1000 instrument is WBC 0.14%, RBC 0.26%, PLT 2.7%, and Hb 1.9%. HCT is 0.69%, and the difference is clinically acceptable. Conclusion It can be considered that the two instruments have good correlation and consistency, and the two instruments can replace each other.
机译:背景通过对2018年山东省外部质量评估中的全血细胞计数结果进行分析,以比较迈瑞BC-5180和Sysmex XN-1000仪器。方法在外部质量评估中,整个过程中共发出了10批质量控制材料年。测试项目为WBC,RBC,Hb,PLT和HCT。使用Mindray BC-5180和Sysmex XN-1000的实验室进行了筛选,并通过t检验,Passing-Bablok回归分析和Bland-Altman分析对结果进行了分析。结果筛选了使用Mindray BC-5180仪器的36个实验室和使用Sysmex XN-1000仪器的36个实验室,并且两个仪器结果之间的平均差异不显着(P>?0.05,t检验)。 Passing-Bablok回归分析显示,每个测试项目的回归方程拦截的95%置信区间包含0,斜率的95%置信区间包含1,r>?0.98,表明相关性良好。 Bland-Altman分析显示,两种仪器在95%的一致性极限内(WBC97.2%,RBC95.6%,PLT97.2%,Hb96.7%,HCT97.5%)都具有超过95%的点。在一致性极限内,Mindray BC-5180仪器与Sysmex XN-1000仪器之间的差的绝对值为WBC 0.14%,RBC 0.26%,PLT 2.7%和Hb 1.9%。 HCT为0.69%,差异在临床上是可以接受的。结论可以认为这两种工具具有良好的相关性和一致性,并且可以相互替代。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号