Precision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose Monitoring System

机译:OneTouch Select简单血糖监测系统的精度,准确性和用户接受度

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Background: The OneTouch? SelectSimple? blood glucose monitoring system (BGMS) is a device for self-monitoring of blood glucose designed for ease of use. Alarms alert subjects to low [20–69 mg/dl (1.1–3.8 mmol/liter)], high [180–239 mg/dl (9.9–13.2 mmol/liter)], and very high [240–600 mg/dl (13.3–33.1 mmol/liter)] blood glucose readings. Methods: Repeatability in blood and intermediate precision with aqueous controls were examined using blood from one donor adjusted to different glucose concentrations, and tested with 10 meters and 1 test-strip lot. System accuracy was evaluated with blood samples from 100 diabetes patients tested on 3 test-strip lots, compared with a reference system (YSI 2300 STAT). To test user accuracy, patients ( n = 156) and health care professionals (HCPs) tested subject blood with the SelectSimple twice. Health care professionals evaluated subject BGMS technique after a 3–5 day home-testing period. Users evaluated the instructions for use and responded to a user acceptance questionnaire. Results: In repeatability and intermediate precision testing, the SelectSimple BGMS had a coefficient of variation of ≤5% or standard deviation of ≤5 mg/dl. In the clinical accuracy study, 100% of measurements <75 mg/dl (4.2 mmol/liter) were within ±15 mg/dl (0.8 mmol/liter) of reference value, and 99.6% of measurements ≥75 mg/dl (4.2 mmol/liter) were within ±20%. Patients were able to use the BGMS appropriately and evaluated it as easy to use. Acceptance of the SelectSimple BGMS was within predefined limits. Conclusions: In these studies, the SelectSimple BGMS met all criteria for precision, system, and user accuracy, was easy to use, and was well accepted by patients.
机译:背景:OneTouch? SelectSimple?血糖监测系统(BGMS)是一种用于自我血糖监测的设备,旨在易于使用。警报会警告对象低[20–69 mg / dl(1.1–3.8 mmol / l)],高[180–239 mg / dl(9.9–13.2 mmol / l)]和非常高[240–600 mg / dl (13.3–33.1 mmol / L)]血糖读数。方法:使用来自一位供体的血液,将其调节至不同的葡萄糖浓度,检查血液的重复性和中等精度,并使用10米和1条测试条进行测试。与参考系统(YSI 2300 STAT)相比,使用3个测试条测试了100位糖尿病患者的血液样本来评估系统的准确性。为了测试用户的准确性,患者(n = 156)和医疗保健专业人员(HCP)使用SelectSimple对受试者血液进行了两次测试。医护专业人员经过3至5天的家庭测试期后,评估了受试者的BGMS技术。用户评估了使用说明,并回复了用户接受调查表。结果:在重复性和中等精度测试中,SelectSimple BGMS的变异系数≤5%或标准偏差≤5mg / dl。在临床准确性研究中,<75 mg / dl(4.2 mmol / L)的测量值的100%在参考值的±15 mg / dl(0.8 mmol / L)的范围内,≥75mg / dl(4.2的测量值)的99.6%摩尔/升)在±20%以内。患者能够适当地使用BGMS,并对其使用进行了评估。 SelectSimple BGMS的接受程度在预定义的范围内。结论:在这些研究中,SelectSimple BGMS满足了精度,系统和用户准确性的所有标准,易于使用,并被患者广泛接受。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号