Impact of the various kVp settings used during computed tomography (CT) simulation that provides data for heterogeneity corrected dose distribution calculations in patients undergoing external beam radiotherapy with either high‐energy photon or electron beams have been investigated. The change of the Hounsfield Unit (HU) values due to the influence of kVp settings and geometrical distribution of various tissue substitute materials has also been studied. The impact of various kVp settings and electron density (ED) distribution on the accuracy of dose calculation in high‐energy photon beams was found to be well within 2%. In the case of dose distributions obtained with a commercially available Monte Carlo dose calculation algorithm for electron beams, differences of more than 10% were observed for different geometrical setups and kVp settings. Dose differences for the electron beams are relatively small at shallow depths but increase with depth around lower isodose values. PACS numbers: 87.57.Q‐, 87.55.D‐
机译:研究了在计算机断层扫描(CT)模拟过程中使用的各种kVp设置的影响,该设置为使用高能光子或电子束进行外部束放射治疗的患者提供了异质校正剂量分布计算的数据。还研究了由于kVp设置和各种组织替代材料的几何分布的影响而引起的Hounsfield单位(HU)值的变化。发现各种kVp设置和电子密度(ED)分布对高能光子束剂量计算精度的影响都在2%以内。在使用可商购的电子束蒙特卡洛剂量计算算法获得剂量分布的情况下,对于不同的几何设置和kVp设置,观察到的差异超过10%。电子束的剂量差在浅深度处相对较小,但在较低的等剂量值附近随深度而增加。 PACS编号:87.57.Q-,87.55.D-

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号