Tolerance levels of CT number to electron density table for photon beam in radiotherapy treatment planning system

机译:放射治疗计划系统中光子束CT数对电子密度表的容忍度

获取原文
           

摘要

Abstract The accuracy of computed tomography number to electron density (CT-ED) calibration is a key component for dose calculations in an inhomogeneous medium. In a previous work, it was shown that the tolerance levels of CT-ED calibration became stricter with an increase in tissue thickness and decrease in the effective energy of a photon beam. For the last decade, a low effective energy photon beam (e.g., flattening-filter-free (FFF)) has been used in clinical sites. However, its tolerance level has not been established yet. We established a relative electron density (ED) tolerance level for each tissue type with an FFF beam. The tolerance levels were calculated using the tissue maximum ratio (TMR) and each corresponding maximum tissue thickness. To determine the relative ED tolerance level, TMR data from a Varian accelerator and the adult reference computational phantom data in the International Commission on Radiological Protection publication 110 (ICRP-110 phantom) were used in this study. The 52 tissue components of the ICRP-110 phantom were classified by mass density as five tissues groups including lung, adipose/muscle, cartilage/spongy-bone, cortical bone, and tooth tissue. In addition, the relative ED tolerance level of each tissue group was calculated when the relative dose error to local dose reached 2%. The relative ED tolerances of a 6?MVFFF beam for lung, adipose/muscle, and cartilage/spongy-bone were ±0.044, ±0.022, and ±0.044, respectively. The thicknesses of the cortical bone and tooth groups were too small to define the tolerance levels. Because the tolerance levels of CT-ED calibration are stricter with a decrease in the effective energy of the photon beam, the tolerance levels are determined by the lowest effective energy in useable beams for radiotherapy treatment planning systems.
机译:摘要计算机体层摄影术数对电子密度(CT-ED)校准的准确性是非均匀介质中剂量计算的关键组成部分。在以前的工作中,已经表明,随着组织厚度的增加和光子束有效能量的降低,CT-ED校准的公差级别变得更加严格。在过去的十年中,在临床场所使用了低有效能量的光子束(例如,无展平滤波器的(FFF))。但是,其公差等级尚未确定。我们用FFF束建立了每种组织类型的相对电子密度(ED)耐受水平。使用组织最大比率(TMR)和每个相应的最大组织厚度来计算耐受水平。为了确定相对ED耐受水平,本研究使用了国际放射防护委员会出版物110(ICRP-110幻影)中的Varian加速器的TMR数据和成人参考计算幻象数据。 ICRP-110体模的52个组织成分按质量密度分类为五个组织组,包括肺,脂肪/肌肉,软骨/脊椎,皮质骨和牙齿组织。此外,当相对于局部剂量的相对剂量误差达到2%时,计算每个组织组的相对ED耐受水平。 6?MVFFF光束对肺,脂肪/肌肉和软骨/脊柱骨的相对ED容差分别为±0.044,±0.022和±0.044。皮质骨和牙齿组的厚度太小,无法定义公差水平。由于CT-ED校准的公差等级随着光子束有效能量的降低而变得更加严格,因此公差等级由放射治疗计划系统可用光束中的最低有效能量确定。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号