首页> 外文期刊>HardwareX >Implementation of a flexible, open-source platform for ion mobility spectrometry
【24h】

Implementation of a flexible, open-source platform for ion mobility spectrometry

机译:实施用于离子迁移谱分析的灵活的开源平台

获取原文

摘要

When operated as a stand-alone device, an ion mobility spectrometer (IMS) routinely offers low limits of detection (pptv-range) for gas-phase analytes even for measurement times less than a second. Mass analyzers further enhance the analytical power of IMS separations, however, high performance drift-cell IMS instruments are often highly customized, relatively large, and require extensive expertise to operate. In this work we present an optimized, low cost IMS system that leverages an easy-to-assemble ion gating structure that enables IMS spectra with resolving powers exceeding 90 for a drift cell only 10?cm in length. The IMS presented in this work consists of stacked rings divided by spacers all fabricated from printed circuit boards (PCB). The rings are connected via a slotted PCB-board containing a surface mounted voltage divider that connects directly to the ring electrodes allowing a fast and easy assembly. This highly modular design enables e.g. the realization of variable drift tube lengths or single and dual gate setups. Instead of the commonly used Bradbury Nielsen gates, the IMS is equipped with a 3-grid ion gate allowing the generation of short (<50?μs) ion packets increasing the resolving power of the instrument.
机译:当作为独立设备运行时,离子迁移谱仪(IMS)通常为气相分析物提供低检测限(pptv范围),即使测量时间不到一秒钟。质量分析仪进一步增强了IMS分离的分析能力,但是,高性能漂移池IMS仪器通常高度定制,相对较大,并且需要大量的专业知识才能操作。在这项工作中,我们提出了一种优化的,低成本的IMS系统,该系统利用易于组装的离子门控结构,对于长度仅为10?cm的漂移池,其IMS光谱的分辨力超过90。这项工作中提出的IMS由堆叠的环组成,这些环被垫片分隔,这些垫片都是由印刷电路板(PCB)制造的。环通过开槽的PCB板连接,该PCB板包含一个表面安装的分压器,该分压器直接连接到环电极,可以快速简便地进行组装。这种高度模块化的设计使得例如可变漂移管长度或单闸门和双闸门设置的实现。 IMS代替了常用的Bradbury Nielsen门,而是配备了3格离子门,可产生短的(<50?s)离子包,从而提高了仪器的分辨能力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号