The dedicated murine PET (MuPET) scanner is a high-resolution, high-sensitivity, and low-cost preclinical PET camera designed and manufactured at our laboratory. In this article, we report i'/> Engineering and Performance (NEMA and Animal) of a Lower-Cost Higher-Resolution Animal PET/CT Scanner Using Photomultiplier-Quadrant-Sharing Detectors
首页> 外文期刊>The Journal of Nuclear Medicine >Engineering and Performance (NEMA and Animal) of a Lower-Cost Higher-Resolution Animal PET/CT Scanner Using Photomultiplier-Quadrant-Sharing Detectors
【24h】

Engineering and Performance (NEMA and Animal) of a Lower-Cost Higher-Resolution Animal PET/CT Scanner Using Photomultiplier-Quadrant-Sharing Detectors

机译:使用光电倍增象限共享探测器的低成本高分离度动物PET / CT扫描仪的工程和性能(NEMA和动物)

获取原文
           

摘要

id="p-2">The dedicated murine PET (MuPET) scanner is a high-resolution, high-sensitivity, and low-cost preclinical PET camera designed and manufactured at our laboratory. In this article, we report its performance according to the NU 4-2008 standards of the National Electrical Manufacturers Association (NEMA). We also report the results of additional phantom and mouse studies. >Methods: The MuPET scanner, which is integrated with a CT camera, is based on the photomultiplier-quadrant-sharing concept and comprises 180 blocks of 13 ?— 13 lutetium yttrium oxyorthosilicate crystals (1.24 ?— 1.4 ?— 9.5 mm3) and 210 low-cost 19-mm photomultipliers. The camera has 78 detector rings, with an 11.6-cm axial field of view and a ring diameter of 16.6 cm. We measured the energy resolution, scatter fraction, sensitivity, spatial resolution, and counting rate performance of the scanner. In addition, we scanned the NEMA image-quality phantom, Micro Deluxe and Ultra-Micro Hot Spot phantoms, and 2 healthy mice. >Results: The system average energy resolution was 14% at 511 keV. The average spatial resolution at the center of the field of view was about 1.2 mm, improving to 0.8 mm and remaining below 1.2 mm in the central 6-cm field of view when a resolution-recovery method was used. The absolute sensitivity of the camera was 6.38% for an energy window of 350-650 keV and a coincidence timing window of 3.4 ns. The system scatter fraction was 11.9% for the NEMA mouselike phantom and 28% for the ratlike phantom. The maximum noise-equivalent counting rate was 1,100 at 57 MBq for the mouselike phantom and 352 kcps at 65 MBq for the ratlike phantom. The 1-mm fillable rod was clearly observable using the NEMA image-quality phantom. The images of the Ultra-Micro Hot Spot phantom also showed the 1-mm hot rods. In the mouse studies, both the left and right ventricle walls were clearly observable, as were the Harderian glands. >Conclusion: The MuPET camera has excellent resolution, sensitivity, counting rate, and imaging performance. The data show it is a powerful scanner for preclinical animal study and pharmaceutical development.
机译:id =“ p-2”>专用的鼠类PET(MuPET)扫描仪是由我们的实验室设计和制造的高分辨率,高灵敏度,低成本的临床前PET相机。在本文中,我们根据国家电气制造商协会(NEMA)的NU 4-2008标准报告其性能。我们还报告了其他幻像和鼠标研究的结果。 >方法: MuPET扫描仪与CT相机集成在一起,基于光电倍增-象限共享概念,包括180块13?—13氧化oxy酸钇钇晶体(1.24?-1.4?)。 — 9.5毫米 3 )和210低成本的19毫米光电倍增管。该摄像机具有78个检测器环,轴向视场为11.6厘米,环直径为16.6厘米。我们测量了扫描仪的能量分辨率,散射分数,灵敏度,空间分辨率和计数率性能。此外,我们还扫描了NEMA图像质量体模,Micro Deluxe和Ultra-Micro Hot Spot体模以及2只健康小鼠。 >结果:系统平均能量分辨率为511 keV时为14%。使用分辨率恢复方法时,视场中心的平均空间分辨率约为1.2毫米,提高到0.8毫米,而在中心6厘米视场中的平均空间分辨率保持在1.2毫米以下。对于350-650 keV的能量窗口和3.4 ns的重合定时窗口,摄像机的绝对灵敏度为6.38%。 NEMA小鼠模型的系统散布率为11.9%,大鼠模型为28%。鼠模的最大等效噪声计数率为57 MBq时为1100,鼠模的最大等效噪声计数率为65 MBq时为352 kcps。使用NEMA图像质量幻像可以清楚地观察到1毫米可填充杆。超微型热点体模的图像还显示了1毫米热棒。在小鼠研究中,左心室壁和右心室壁均清晰可见,哈德氏腺也是如此。 >结论:MuPET相机具有出色的分辨率,灵敏度,计数率和成像性能。数据表明,它是用于临床前动物研究和药物开发的强大扫描仪。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号