Consequences of the Patient's Mis-centering on the Radiation Dose and Image Quality in CT Imaging – Phantom and Clinical Study

机译:CT影像中患者以放射剂量和图像质量为中心的后果–幻影和临床研究

获取原文
       

摘要

The position of the patient highly influences the functioning of the ATCM in CT imaging. The effect of different mis-centerings on the CTDIvol dose was determined during PMMA and water phantom simulations, which were also used for noise assessment. The results show that a 50 mm mis-centering (with the phantom placed above the isocenter) can cause an increase of the CTDIvol by 47 % associated with the lower standard deviation of the HU signal, whilst a -50 mm mis-centering (with the phantom placed below the isocenter) leads to a decrease of the CTDIvol by 35% and is associated with a noisier image. A mean value of the mis-centering determined for 473 patients was -43 mm. A total of 470 of 473 patients were mis-centered below the X-ray tube, which shows an inclination of radiographers to place patients below the isocenter. This inclination was more significant for smaller patients.
机译:患者的位置在很大程度上影响ATCM在CT成像中的功能。在PMMA和水模仿真中确定了不同偏心对CTDIvol剂量的影响,这些仿真也用于噪声评估。结果表明,50毫米的偏心度(幻影位于等角点上方)可能导致CTDIvol增加47%,这与HU信号的标准差较低有关,而-50毫米的偏心度(位于等角点下方的体模)会导致CTDIvol降低35%,并且与噪点图像相关。为473例患者确定的偏心平均值为-43毫米。 473位患者中有470位位于X射线管下方偏心,这表明放射线照相师倾向于将患者置于等中心线以下。对于较小的患者,这种倾向更为明显。