The correlation between the first heart sound and cardiac output as measured by using digital esophageal stethoscope under anaesthesia

机译:麻醉下使用数字食管听诊器测量的第一心音与心输出量之间的相关性

获取原文
       

摘要

Objective: The use of an esophageal stethoscope is a basic heart sounds monitoring procedure performed in patients under general anesthesia. As the size of the first heart sound can express the left ventricle function, its correlation with cardiac output should be investigated. The aim of this study was to investigate the effects of cardiac output (CO) on the first heart sound (S1) amplitude.Methods: Six male beagles were chosen. The S1 was obtained with the newly developed esophageal stethoscope system. CO was measured using NICOM, a non-invasive CO measuring device. Ephedrine and beta blockers were administered to the subjects to compare changes in figures, and the change from using an inhalation anesthetic was also compared.Results: The S1 amplitude displayed positive correlation with the change rate of CO (r = 0.935, p < 0.001). The heart rate measured using the esophageal stethoscope and ECG showed considerably close figures through the Bland-Altman plot and showed a high positive correlation (r = 0.988, p < 0.001).Conclusion: In beagles, the amplitude of S1 had a significant correlation with changes in CO in a variety of situations.doi: http://dx.doi.org/10.12669/pjms.302.4383How to cite this:Shin YD, Yim KH, Park SH, Jeon YW, Bae JH, Lee TS, et al. The correlation between the first heart sound and cardiac output as measured by using digital esophageal stethoscope under anaesthesia. Pak J Med Sci 2014;30(2):276-281. doi: http://dx.doi.org/10.12669/pjms.302.4383This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
机译:目的:食管听诊器的使用是在全身麻醉下对患者进行的基本心音监测程序。由于第一个心音的大小可以表达左心室功能,因此应研究其与心输出量的相关性。这项研究的目的是调查心输出量(CO)对第一个心音(S1)振幅的影响。方法:选择六只雄性比格犬。 S1是通过新开发的食管听诊器系统获得的。使用非侵入式CO测量设备NICOM测量CO。向受试者服用麻黄碱和β受体阻滞剂以比较数字变化,还比较吸入麻醉药的变化。结果:S1振幅与CO的变化率呈正相关(r = 0.935,p <0.001) 。食管听诊器和ECG测得的心率在Bland-Altman图上显示出非常接近的数字,并显示出很高的正相关性(r = 0.988,p <0.001)。结论:在小猎犬中,S1的振幅与在各种情况下CO的变化。doi:http://dx.doi.org/10.12669/pjms.302.4383如何引用这一点:申YD,Yim KH,Park SH,Jeon YW,Bae JH,Lee TS等等在麻醉下使用数字食管听诊器测量的第一心音与心输出量之间的相关性。 Pak J Med Sci 2014; 30(2):276-281。 doi:http://dx.doi.org/10.12669/pjms.302.4383。这是根据知识共享署名许可(http://creativecommons.org/licenses/by/3.0)的条款分发的开放获取文章,只要适当引用了原始作品,便可以在任何介质中不受限制地使用,分发和复制。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号