Objective: To evaluate the performance of Sysmex KX – 21 for the estimation of Platelet and WBC count in comparison with microscopy. Study Design: It is prospective cohort study. Place and Duration: It was conducted in the Patho-logy department, Allama Iqbal Medical College, from 7th Oct. 2009 to 6th Nov. 2009. Materials and Methods: Out of the total 1500 samp-les, 1458 (97%) samples after exclusions were finally evaluated for the study. On each sample, automated blood cell count was performed using Sysmex KX – 21 and the results were compared by peripheral smear examination for Platelet count, WBC count and its dif-ferential. The samples were from Medicine, Pediatrics and Oncology wards only. Results: Out of 1458 samples, WBC count of 1452 (99.5%) samples matched the total count on peripheral smear examination. 06 (0.5%) samples showed spur-ious WBC count due to the presence of nucleated
机译:目的:与显微镜相比,评估Sysmex KX – 21在评估血小板和白细胞计数方面的性能。研究设计:这是前瞻性队列研究。地点和时间:从2009年10月7日至2009年11月6日在阿拉玛·伊克巴尔医学院的病理学系进行。材料与方法:在1500个样本中,抽取1458个样本(97%)。最后对排除进行评估以进行研究。在每个样品上,使用Sysmex KX – 21进行自动血细胞计数,并通过外周涂片检查对血小板计数,WBC计数及其差异进行比较。样本仅来自医学,儿科和肿瘤病房。结果:在1458个样本中,有1452个样本的WBC计数(99.5%)与外周涂片检查的总数相符。 06(0.5%)样本显示有核的白细胞计数