Comparison of 265 nm Femtosecond and 213 nm Nanosecond Laser Ablation Inductively Coupled Plasma Mass Spectrometry for Pb Isotope Ratio Measurements

机译:265 nm飞秒和213 nm纳秒激光烧蚀电感耦合等离子体质谱法用于Pb同位素比测量的比较

获取原文
   

获取外文期刊封面封底 >>

       

摘要

The analytical performance of 265 nm femtosecond laser ablation (fs-LA) and 213 nm nanosecond laser ablation (ns-LA) systems coupled with multi-collector inductively coupled plasma mass spectrometry (MC-ICPMS) for Pb isotope ratio measurements of solder were compared. Although the time-resolved signals of Pb measured by fs-LA-MC-ICPMS showed smoother signals compared to those obtained by ns-LA-MC-ICPMS, similar precisions on Pb isotope ratio measurements were obtained between them, even though their operating conditions were slightly different. The mass bias correction of the Pb isotope ratio measurement was carried out by a comparison method using a Pb standard solution prepared from NIST SRM 981 Pb metal isotopic standard, which was introduced into the ICP by a desolvation nebulizer (DSN) via a dual-sample introduction system, and it was successfully demonstrated for Pb isotope ratio measurements for either NIST 981 metal isotopic standard or solder by fs-LA-MC-ICPMS since the analytical results agreed well with the certified value as well as the determined value within their standard deviations obtained and the expanded uncertainty of the certified or determined value. The Pb isotope ratios of solder obtained by ns-LA-MC-ICPMS also showed agreement with respect to the determined value within their standard deviations and expanded uncertainty. From these results, it was evaluated that the mass bias correction applied in the present study was useful and both LA-MC-ICPMS could show similar analytical performance for the Pb isotope ratio microanalysis of metallic samples such as solder.
机译:比较了265 nm飞秒激光烧蚀(fs-LA)和213 nm纳秒激光烧蚀(ns-LA)系统与多收集器电感耦合等离子体质谱仪(MC-ICPMS)的焊料中Pb同位素比测量的分析性能。尽管与通过ns-LA-MC-ICPMS获得的信号相比,通过fs-LA-MC-ICPMS测量得到的Pb的时间分辨信号显示出更平滑的信号,但即使在它们的工作条件下,它们之间的Pb同位素比测量也获得了相似的精度略有不同。使用NIST SRM 981 Pb同位素标准品制备的Pb标准溶液通过比较方法进行Pb同位素比测量的质量偏差校正,该溶液通过去溶剂雾化器(DSN)通过双样品引入ICP引入系统,并通过fs-LA-MC-ICPMS成功地证明了NIST 981金属同位素标准品或焊料的Pb同位素比测量,因为分析结果与认证值以及标准偏差内的测定值非常吻合获得认证或确定值的不确定性。通过ns-LA-MC-ICPMS获得的焊料的Pb同位素比在其标准偏差和扩展不确定性范围内也与测定值一致。从这些结果可以看出,本研究中应用的质量偏差校正是有用的,并且两种LA-MC-ICPMS都可以对金属样品(如焊料)的Pb同位素比微分析显示出相似的分析性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号