首页> 外文期刊>Analytical methods >Repeatability and reproducibility of desorption electrospray ionization-mass spectrometry (DESI-MS) for the imaging analysis of human cancer tissue: a gateway for clinical applications
【24h】

Repeatability and reproducibility of desorption electrospray ionization-mass spectrometry (DESI-MS) for the imaging analysis of human cancer tissue: a gateway for clinical applications

机译:解吸电喷雾电离质谱(DESI-MS)用于人类癌症组织成像分析的可重复性和重现性:临床应用的门户

获取原文
           

摘要

In this study, we aim to demonstrate the repeatability and reproducibility of DESI-MS for the imaging analysis of human cancer tissue using a set of optimal geometric and electrospray solvent parameters. Oesophageal cancer tissue was retrieved from four quadrants of a freshly removed tumor specimen, snap frozen, cryo-sectioned and mounted on glass slides for DESI-MS image acquisition. Prior to assessing precision, optimal geometric and electrospray solvent parameters were determined to maximize the number of detected lipid species and associated Total Ion Count (TIC). The same settings were utilized for all subsequent experiments. Repeatability measurements were performed using the same instrument, by the same operator on a total of 16 tissue sections (four from each quadrant of the tumor). Reproducibility measurements were determined in a different laboratory, on a separate DESI-MS platform and by an independent operator on 4 sections of one quadrant and compared to the corresponding measurements made for the repeatability experiments. The mean ± SD CV of lipid ion intensities was found to be 22 ± 7% and 18 ± 8% as measures of repeatability and reproducibility, respectively. In conclusion, DESI-MS has acceptable levels of reproducibility for the analysis of lipids in human cancer tissue and is suitable for the purposes of clinical research and diagnostics...
机译:在这项研究中,我们旨在证明DESI-MS使用一组最佳几何和电喷雾溶剂参数对人类癌症组织进行成像分析的可重复性和重现性。从新鲜取出的肿瘤标本的四个象限中取出食道癌组织,快速冷冻,冷冻切片并安装在载玻片上以进行DESI-MS图像采集。在评估精度之前,确定最佳的几何和电喷雾溶剂参数,以使检测到的脂质种类和相关的总离子数(TIC)最大化。所有后续实验均使用相同的设置。使用相同的仪器,由相同的操作员对总共16个组织切片(肿瘤的每个象限中有四个)进行重复性测量。在不同的实验室中,在单独的DESI-MS平台上以及由独立的操作员在一个象限的4个部分上确定可重复性测量,并将其与为重复性实验进行的相应测量进行比较。发现脂质离子强度的平均值±SD CV分别为22±7%和18±8%,作为重复性和再现性的量度。总而言之,DESI-MS具有可接受水平的重现性,可用于分析人类癌症组织中的脂质,适用于临床研究和诊断的目的...

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号