Background and purpose The most frequently used bones for mechanical testing of orthopedic and trauma devices are fresh frozen cadaveric bones, embalmed cadaveric bones, and artificial composite bones. Even today, the comparability of these different bone types has not been established. Methods We tested fresh frozen and embalmed cadaveric femora that were similar concerning age, sex, bone mineral density, and stiffness. Artificial composite femora were used as a reference group. Testing parameters were pullout forces of cortex and cancellous screws, maximum load until failure, and type of fracture generated. Results Stiffness and type of fracture generated (Pauwels III) were similar for all 3 bone types (fresh frozen: 969 N/mm, 95% confidence interval (CI): 897–1,039; embalmed: 999 N/mm, CI: 875–1,121; composite: 946 N/mm, CI: 852–1,040). Furthermore, no significant differences were found between fresh frozen and embalmed femora concerning pullout forces of cancellous screws (fresh frozen: 654 N, CI: 471...
机译:背景和目的用于骨科和创伤器械的机械测试的最常用骨头是新鲜冷冻的尸体骨头,防腐的尸体骨头和人造复合骨。直到今天,这些不同骨骼类型的可比性尚未建立。方法我们测试了新鲜冷冻和防腐的尸体股骨,它们在年龄,性别,骨矿物质密度和刚度方面相似。人工复合股骨被用作参考组。测试参数为皮质和松质骨螺钉的拔出力,直至破裂的最大载荷以及产生的骨折类型。结果所有三种骨类型(刚冻:969 N / mm,95%置信区间(CI):897–1,039;防腐剂:999 N / mm,CI:875–)产生的刚度和骨折类型(Pauwels III)相似1,121;复合材料:946 N / mm,CI:852–1,040)。此外,新鲜的冷冻和防腐的股骨之间在松质螺钉的拔出力方面没有发现显着差异(新鲜冷冻:654 N,CI:471 ...

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号