Polyethylene acetabular wear in hip prostheses: Computer-simulated quantification of error caused by changes in pelvic orientation and direction of wear

机译:髋关节假体中的聚乙烯髋臼磨损:计算机模拟的由骨盆方向和磨损方向变化引起的误差的量化

获取原文
           

摘要

Background and purpose ?Polyethylene is commonly employed for bearings in acetabular cups used in hip replacements. Assessment of in vivo wear is important for evaluation and monitoring of wear in individual patients, as well as in different implant designs. Polyethylene wear is quantified by comparisons of radiographic measurements made on sequential pelvic radiographs. Variations in pelvic orientation and variation in direction of wear may cause underestimation of polyethylene wear measurements. The purpose of this study was to quantify these effects on 2-dimensional measurements of polyethylene wear. Methods ?A computer program designed to simulate radiographs was employed to generate virtual radiographs of a virtual pelvis with a total hip replacement. Effects caused by variation in pelvic spatial orientation and variations in wear direction on wear measurements were analyzed separately. A Monte Carlo computational algorithm was employed to describe the combined effects of these two factors. Results ?Variation in pelvic orientation induced a mean underestimation of wear of 0.4% (0–2.6). Variation in direction of wear introduced a mean underestimation of 8.5% (0–42). A mean underestimation of wear of 9% (0–99) was found when varying both pelvic orientation and direction of wear simultaneously. Interpretation ?Errors caused by variations in pelvic orientation and wear direction are likely to be small compared to other sources of error when performing polyethylene wear measurements in acetabular components.
机译:背景和目的?聚乙烯通常用于髋关节置换用髋臼杯中的轴承。体内磨损的评估对于评估和监测单个患者以及不同植入物设计的磨损非常重要。聚乙烯磨损是通过比较连续盆腔X射线照片上的X射线照片来量化的。骨盆方向的变化和磨损方向的变化可能会导致聚乙烯磨损测量值被低估。这项研究的目的是量化这些对聚乙烯磨损的二维测量的影响。方法:采用设计用于模拟X光片的计算机程序来生成虚拟的骨盆的X光片,并进行全髋关节置换。分别分析了骨盆空间方向变化和磨损方向变化对磨损测量结果的影响。采用蒙特卡洛计算算法来描述这两个因素的综合作用。结果?骨盆方向变化导致平均低估了0.4%(0–2.6)的磨损。磨损方向的变化导致平均低估了8.5%(0-42)。当同时改变骨盆方向和磨损方向时,平均磨损被低估了9%(0–99)。解释?在髋臼组件中进行聚乙烯磨损测量时,与其他误差源相比,由骨盆方向和磨损方向变化引起的误差可能较小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号